A 32-Channel Head Coil Array with Circularly Symmetric Geometry for Accelerated Human Brain Imaging
نویسندگان
چکیده
The goal of this study is to optimize a 32-channel head coil array for accelerated 3T human brain proton MRI using either a Cartesian or a radial k-space trajectory. Coils had curved trapezoidal shapes and were arranged in a circular symmetry (CS) geometry. Coils were optimally overlapped to reduce mutual inductance. Low-noise pre-amplifiers were used to further decouple between coils. The SNR and noise amplification in accelerated imaging were compared to results from a head coil array with a soccer-ball (SB) geometry. The maximal SNR in the CS array was about 120% (1070 vs. 892) and 62% (303 vs. 488) of the SB array at the periphery and the center of the FOV on a transverse plane, respectively. In one-dimensional 4-fold acceleration, the CS array has higher averaged SNR than the SB array across the whole FOV. Compared to the SB array, the CS array has a smaller g-factor at head periphery in all accelerated acquisitions. Reconstructed images using a radial k-space trajectory show that the CS array has a smaller error than the SB array in 2- to 5-fold accelerations.
منابع مشابه
A 32-channel Receive Array Coil for Pediatric Brain Imaging at 3T
Introduction: Pediatric MRI currently requires general anesthesia to preclude motion artifacts. With highly accelerated image encoding in conjunction motion navigators it may be possible to significantly reduce or even eliminate anesthesia. For research studies of brain development where anesthesia is not possible, improved motion robustness can significantly improve the success rate. Highly pa...
متن کامل96-Channel receive-only head coil for 3 Tesla: design optimization and evaluation.
The benefits and challenges of highly parallel array coils for head imaging were investigated through the development of a 3T receive-only phased-array head coil with 96 receive elements constructed on a close-fitting helmet-shaped former. We evaluated several designs for the coil elements and matching circuitry, with particular attention to sources of signal-to-noise ratio (SNR) loss, includin...
متن کاملSize-optimized 32-channel brain arrays for 3 T pediatric imaging.
Size-optimized 32-channel receive array coils were developed for five age groups, neonates, 6 months old, 1 year old, 4 years old, and 7 years old, and evaluated for pediatric brain imaging. The array consisted of overlapping circular surface coils laid out on a close-fitting coil-former. The two-section coil former design was obtained from surface contours of aligned three-dimensional MRI scan...
متن کاملComparison of a 32-channel with a 12-channel head coil: are there relevant improvements for functional imaging?
PURPOSE To evaluate the suitability of a 12- or 32-channel head coil and of a prescan normalization filter for functional magnetic resonance imaging (fMRI) studies at different brain regions. MATERIALS AND METHODS fMRI was obtained from 36 volunteers executing a visually instructed motor paradigm using a 12-channel head matrix coil and a 32-channel phased-array head coil with and without pres...
متن کاملA Specialized Multi-Transmit Head Coil for High Resolution fMRI of the Human Visual Cortex at 7T
PURPOSE To design, construct and validate radiofrequency (RF) transmit and receive phased array coils for high-resolution visual cortex imaging at 7 Tesla. METHODS A 4 channel transmit and 16 channel receive array was constructed on a conformal polycarbonate former. Transmit field efficiency and homogeneity were simulated and validated, along with the Specific Absorption Rate, using [Formula:...
متن کامل