We report on the use of the EGS4/BEAM Monte Carlo technique to predict the output factors for clinically relevant, irregularly shaped inserts as they intercept a linear accelerator's electron beams. The output factor for a particular combination--energy, cone, insert, and source-to-surface distance (SSD)--is defined in accordance with AAPM TG-25 as the product of cone correction factor and insert correction factor, evaluated at the depth of maximum dose. Since cone correction factors are easily obtained, we focus our investigation on the insert correction factors (ICFs). An analysis of the inserts used in routine clinical practice resulted in the identification of a set of seven "idealized" shapes characterized by specific parameters. The ICFs for these shapes were calculated using a Monte Carlo method (EGS4/BEAM) and measured for a subset of them using an ion chamber and well-established measurement methods. Analytical models were developed to predict the Monte Carlo-calculated ICF values for various electron energies, cone sizes, shapes, and SSDs. The goodness-of-fit between predicted and Monte Carlo-calculated ICF values was tested using the Kolmogorov-Smirnoff statistical test. Results show that Monte Carlo-calculated ICFs match the measured values within 2.0% for most of the shapes considered, except for few highly elongated fields, where deviations up to 4.0% were recorded. Predicted values based on analytical modeling agree with measured ICF values within 2% to 3% for all configurations. We conclude that the predicted ICF values based on modeling of Monte Carlo-calculated values could be introduced in clinical use.
منابع مشابه
Small photon field dosimetry using EBT2 Gafchromic film and Monte Carlo simulation
Background: Small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (IMRT) and stereotactic radiosurgery (SRS) treatments. Accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. Hence the EBT2 (external beam therapy) Gafchromic film was used for dosimetry of sm...
متن کاملDose Calculations for Lung Inhomogeneity in High-Energy Photon Beams and Small Beamlets: A Comparison between XiO and TiGRT Treatment Planning Systems and MCNPX Monte Carlo Code
Introduction Radiotherapy with small fields is used widely in newly developed techniques. Additionally, dose calculation accuracy of treatment planning systems in small fields plays a crucial role in treatment outcome. In the present study, dose calculation accuracy of two commercial treatment planning systems was evaluated against Monte Carlo method. Materials and Methods Siemens Once or linea...
متن کاملA Monte Carlo simulation and dosimetric verification of physical wedges used in radiation therapy
Background: The presence of a wedge filter in the beam trajectory can modify the beam quality and cause some changes in the dosimetry parameters which are usually difficult to be measured directly and accurately. Material and Methods: In this study the MCNP-4C Monte Carlo code was used to simulate the 9 MV photon beam generated by a linear accelerator. Upon getting a good agreement bet...
متن کاملExact diagonalization and Quantum Monte Carlo studies of quantum dots
We report exact diagonalization and diffusion quantum Monte Carlo calculations of quantum dots (QDs) in which energetics due to electron-electron interactions, magnetic field, and geometrical factors compete and induce interesting ground state configurations. The geometrical effects are generated by different confining potentials such as parabolic shaped, ring shaped, and disordered potentials....
متن کاملMonte Carlo simulation for Neptun 10 PC medical linear accelerator and calculations of output factor for electron beam.
AIM Exact knowledge of dosimetric parameters is an essential pre-requisite of an effective treatment in radiotherapy. In order to fulfill this consideration, different techniques have been used, one of which is Monte Carlo simulation. MATERIALS AND METHODS This study used the MCNP-4C to simulate electron beams from Neptun 10 PC medical linear accelerator. Output factors for 6, 8 and 10 MeV el...