Characterization of an in-vacuum PILATUS 1M detector.
نویسندگان
چکیده
A dedicated in-vacuum X-ray detector based on the hybrid pixel PILATUS 1M detector has been installed at the four-crystal monochromator beamline of the PTB at the electron storage ring BESSY II in Berlin, Germany. Owing to its windowless operation, the detector can be used in the entire photon energy range of the beamline from 10 keV down to 1.75 keV for small-angle X-ray scattering (SAXS) experiments and anomalous SAXS at absorption edges of light elements. The radiometric and geometric properties of the detector such as quantum efficiency, pixel pitch and module alignment have been determined with low uncertainties. The first grazing-incidence SAXS results demonstrate the superior resolution in momentum transfer achievable at low photon energies.
منابع مشابه
PILATUS 2M A Detector for Small Angle X-ray Scattering
For the (coherent) small-angle X-ray scattering (SAXS) beamline X12SA of the Swiss Light Source at the Paul Scherrer Institut, a novel X-ray detector is built, the PILATUS 2M. Contributions in module fabrication, detector assembly, characterization and calibration are covered within the framework of this thesis. Furthermore, data analysis of a particular SAXS experiment demonstrating the perfor...
متن کاملNext Generation X-ray Detectors for In-house Xrd
A novel type X-ray detector, called PILATUS, has been developed at the Paul Scherrer Institut (PSI) in Switzerland during the last decade. PILATUS detectors are two-dimensional hybrid pixel array detectors, which operate in single-photon counting mode. PILATUS detectors feature a very wide dynamic range (1:1,000,000), very short read out time (< 3.0 msec), no readout noise, and very high counti...
متن کاملPicosecond time-resolved laser pump/X-ray probe experiments using a gated single-photon-counting area detector.
The recent developments in X-ray detectors have opened new possibilities in the area of time-resolved pump/probe X-ray experiments; this article presents the novel use of a PILATUS detector to achieve X-ray pulse duration limited time-resolution at the Advanced Photon Source (APS), USA. The capability of the gated PILATUS detector to selectively detect the signal from a given X-ray pulse in 24 ...
متن کاملEIGER detector: application in macromolecular crystallography
The development of single-photon-counting detectors, such as the PILATUS, has been a major recent breakthrough in macromolecular crystallography, enabling noise-free detection and novel data-acquisition modes. The new EIGER detector features a pixel size of 75 × 75 µm, frame rates of up to 3000 Hz and a dead time as low as 3.8 µs. An EIGER 1M and EIGER 16M were tested on Swiss Light Source beam...
متن کاملA Message from the Director Steven Ealick
In May 2009, NE-CAT applied for a high end equipment grant from NCRR to fund the purchase of a Pixel Array Detector (PAD). As a class, PADs are generally acknowledged as the near-term successor to CCD’s at 3rd generation synchrotron crystallography beamlines. The PILATUS-6M, from Dectris Ltd., is the first commercially available PAD suitable for synchrotron-based macromolecular crystallographic...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of synchrotron radiation
دوره 21 Pt 3 شماره
صفحات -
تاریخ انتشار 2014