A clinical trial evaluating the laryngeal mask airway-Supreme in obese children during general anesthesia

نویسندگان

  • Yue Tian
  • Xiu-ying Wu
  • Lu Li
  • Ling Ma
  • Yun-feng Li
چکیده

INTRODUCTION The laryngeal mask airway (LMA)-Supreme is a disposable double-lumen laryngeal mask airway that is widely used in clinical practice. However, its use in obese children has not been evaluated. The aim of this study was to determine whether the LMA-Supreme could perform equally as well as endotracheal intubation in obese children having a minor surgical procedure. MATERIAL AND METHODS After ethical board approval, 100 obese male children receiving non-emergent appendectomy for chronic appendicitis or surgery to correct concealed penis were randomly divided into an endotracheal intubation group and an LMA-Supreme group. Endotracheal intubation was performed under direct vision laryngoscopy. In the LMA group, a size-3 LMA-Supreme was placed and a stomach tube inserted via the drainage tube of the mask. Cardiovascular and respiratory parameters, time taken for placement, placement attempts, time to removal of the endotracheal tube/LMA, length of stay in the post-anesthesia care unit (PACU), and complications were recorded. RESULTS Insertion time was significantly longer (p < 0.001) in the LMA-Supreme group than in the endotracheal intubation group. Peak airway pressure was significantly higher, and pulmonary compliance and PACU stay time lower in the LMA-Supreme group. No significant differences between endotracheal intubation and the LMA-Supreme were seen in other parameters, except for a higher incidence of coughing in the endotracheal intubation group. CONCLUSIONS The LMA-Supreme can be easily inserted and effectively used for airway management in obese children undergoing minor surgery.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of the effect of thiopental sodium-midazolam combination on the hemodynamic response between tracheal intubation and laryngeal mask airway insertion: A clinical trial study

Background and Aim: The purpose of this study was to compare the effect of thiopental sodium-midazolam combination on the stress response and hemodynamic changes between tracheal intubation and laryngeal mask airway insertion during airway management. Materials and Methods: In this clinical trial study, 70 patients ASA 1, 2 candidates for elective surgery were randomly divided into two groups o...

متن کامل

Limitation of Laryngeal Mask Airway Usage in Open-Neck Procedure

I read the manuscript entitled “Case Report of An Open Neck Procedure Complication Associated with Laryngeal Mask Airway Use” by Subramaniam T et al. This is a valuable manuscript describing iatrogenic complications during level-2 cervical lymph node biopsy, in which the airway is managed with laryngeal mask airway (LMA). During the procedure, the surgeon realized that the mass was actually the...

متن کامل

Hemodynamic responses to insertion of the laryngeal mask airway versus Combitube

Introduction: One of the most common problems during laryngoscopy involves sympathetic system stimulation and the subsequent hemodynamic changes. It is a key measure to maintain the hemodynamic stability of patients with cardiovascular diseases during induction of anesthesia. This study attempted to compare the hemodynamic effects of laryngeal mask airway and combitube. Methods: This prospec...

متن کامل

Comparing the Effect of Laryngeal Mask Airway And Endotracheal Tube Insertion on Intraocular Pressure in Cataract Surgery

Background and purpose: The method of anesthesia affects intraocular pressure (IOP), especially general anesthesia which is sometimes administered with endotracheal tube (ETT) insertion. The aim of this research was to compare the effect of laryngeal mask airway (LMA) and ETT on intraocular pressure in cataract surgery. Materials and methods: In a single blind randomized clinical trial, 61 pat...

متن کامل

A comparison of i-gel™ and Laryngeal Mask Airway Supreme™ during general anesthesia in infants

Background The i-gel™ (i-gel) and Laryngeal Mask Airway Supreme™ (LMA Supreme) have been safely used in children. We compared the airway performance of the i-gel and LMA Supreme in infants undergoing general anesthesia. Methods Sixty infants with American Society of Anesthesiologists physical status I or II were randomly assigned to place either the i-gel or the LMA Supreme. The size 1 or 1.5...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 13  شماره 

صفحات  -

تاریخ انتشار 2017