The significance of radiolucent zones surrounding pedicle screws. Definition of screw loosening in spinal instrumentation.
نویسندگان
چکیده
We examined the radiographs from a prospective clinical study of fixation by pedicle screws and those from an experimental study in a sheep model. In the clinical study, instruments were removed from 21 patients after implantation for 11 to 16 months and the extraction torques of the screws were recorded. A structured protocol was used for the radiological examinations. In the experimental study, loaded pedicle screw instrumentations were implanted in the sheep for six or 12 weeks. After radiological examination the pull-out resistance and the histological characteristics were studied. In the clinical study, all screws with radiolucent zones had a significantly reduced mean extraction torque compared with screws without radiolucent zones (16 +/- 10 Ncm v 403 +/- 220 Ncm; p < 0.0001). In the experimental study the mean maximum pull-out resistance for the screws with radiolucent zones was significantly lower than for those with no radiolucency (243 +/- 156 N v2214 +/- 578 N; p = 0.0006) and the mean bone-to-screw contact was reduced for screws with zones compared with those without zones (8 +/- 9% v 55 +/- 29%; p = 0.0002). Our findings showed that all screws with radiolucent zones had low extraction torques or low pull-out resistance. A radiolucent zone is a good indicator of loosening of a pedicle screw.
منابع مشابه
Intraoperative biomechanics of lumbar pedicle screw loosening following successful arthrodesis.
Pedicle screw loosening has been implicated in recurrent back pain after lumbar spinal fusion, but the degree of loosening has not been systematically quantified in patients. Instrumentation removal is an option for patients with successful arthrodesis, but remains controversial. Here, we quantified pedicle screw loosening by measuring screw insertion and/or removal torque at high statistical p...
متن کاملHydroxyapatite coating improves fixation of pedicle screws. A clinical study.
We investigated the effects of hydroxyapatite (HA) coating on the purchase of pedicle screws. A total of 23 consecutive patients undergoing lumbar fusion was randomly assigned to one of three treatment groups. The first received uncoated stainless-steel screws, the second screws which were partly coated with HA, and the third screws which were fully coated. The insertion torque was recorded. Af...
متن کاملPedicle screw surface coatings improve fixation in nonfusion spinal constructs.
STUDY DESIGN Biomechanical and histologic analysis. OBJECTIVE To compare the strength of the bone-screw interface of standard uncoated pedicle screws with screws treated with hydroxyapatite (HA), titanium plasma spray (TPS), and a composite HA-TPS coating. SUMMARY OF BACKGROUND DATA Transpedicular screw fixation has become the gold standard in the treatment of various thoracolumbar spinal c...
متن کاملPullout Strength after Expandable Polymethylmethacrylate Transpedicular Screw Augmentation for Pedicle Screw Loosening
OBJECTIVE Pedicle screw fixation for spine arthrodesis is a useful procedure for the treatment of spinal disorders. However, instrument failure often occurs, and pedicle screw loosening is the initial step of a range of complications. The authors recently used a modified transpedicular polymethylmethacrylate (PMMA) screw augmentation technique to overcome pedicle screw loosening. Here, they rep...
متن کاملشکست وسایل تثبیت ستون فقرات
This was a retrospective review of a consecutive series of patients with spinal disease in year 2000, who underwent posterior fusion and instrumentation with Harrington distraction and Cotrel-Dobousset system to evaluate causes of hardware failure. Many cases of clinical failure has been observed in spinal instrumentation used in spinal disorder like spondylolisthesis, fractures, deformities, …...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Journal of bone and joint surgery. British volume
دوره 86 3 شماره
صفحات -
تاریخ انتشار 2004