Relative efficacy of monophasic and biphasic waveforms for transthoracic defibrillation after short and long durations of ventricular fibrillation.

نویسندگان

  • G P Walcott
  • S B Melnick
  • F W Chapman
  • J L Jones
  • W M Smith
  • R E Ideker
چکیده

BACKGROUND Recently, interest has arisen in using biphasic waveforms for external defibrillation. Little work has been done, however, in measuring transthoracic defibrillation efficacy after long periods of ventricular fibrillation. In protocol 1, we compared the efficacy of a quasi-sinusoidal biphasic waveform (QSBW), a truncated exponential biphasic waveform (TEBW), and a critically damped sinusoidal monophasic waveform (CDSMW) after 15 seconds of fibrillation. In protocol 2, we compared the efficacy of the more efficacious biphasic waveform from protocol 1, QSBW, with CDSMW after 15 seconds and 5 minutes of fibrillation. METHODS AND RESULTS In protocol 1, 50% success levels, ED50, were measured after 15 seconds of fibrillation for the 3 waveforms in 6 dogs. In protocol 2, defibrillation thresholds were measured for QSBW and CDSMW after 15 seconds of fibrillation and after 3 minutes of unsupported fibrillation followed by 2 minutes of fibrillation with femoral-femoral cross-circulation. In protocol 1, QSBW had a lower ED50, 16.0+/-4.9 J, than TEBW, 20.3+/-4.4 J, or CDSMW, 27.4+/-6.0 J. In protocol 2, QSBW had a lower defibrillation threshold after 15 seconds, 38+/-10 J, and after 5 minutes, 41.5+/-5 J, than CDSMW after 15 seconds, 54+/-19 J, and 5 minutes, 80+/-30 J, of fibrillation. The defibrillation threshold remained statistically the same for QSBW for the 2 fibrillation durations but rose significantly for CDSMW. CONCLUSIONS In this animal model of sudden death and resuscitation, these 2 biphasic waveforms are more efficacious than the CDSMW at short durations of fibrillation. Furthermore, the QSBW is even more efficacious than the CDSMW at longer durations of fibrillation.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of a novel rectilinear biphasic waveform with a damped sine wave monophasic waveform for transthoracic ventricular defibrillation. ZOLL Investigators.

OBJECTIVES We compared the efficacy of a novel rectilinear biphasic waveform, consisting of a constant current first phase, with a damped sine wave monophasic waveform during transthoracic defibrillation. BACKGROUND Multiple studies have shown that for endocardial defibrillation, biphasic waveforms have a greater efficacy than monophasic waveforms. More recently, a 130-J truncated exponential...

متن کامل

The Role of Biphasic Shocks for Transthoracic Cardioversion of Atrial Fibrillation

The modern generation of transthoracic defibrillators now employ impedance compensated biphasic waveforms. These new devices are superior to those with monophasic waveforms and practice is currently switching to biphasic defibrillators for the treatment of both ventricular and atrial fibrillation. However, there is no universal guideline for the use of biphasic defibrillators in direct current ...

متن کامل

Multicenter comparison of truncated biphasic shocks and standard damped sine wave monophasic shocks for transthoracic ventricular defibrillation. Transthoracic Investigators.

BACKGROUND The most important factor for improving out-of-hospital ventricular fibrillation survival rates is early defibrillation. This can be achieved if small, lightweight, inexpensive automatic external defibrillators are widely disseminated. Because automatic external defibrillator size and cost are directly affected by defibrillation waveform shape and because of the favorable experience ...

متن کامل

Simultaneous Comparison of Many Triphasic Defibrillation Waveforms

Biphasic defibrillation waveforms are now accepted as being more effective at terminating ventricular fibrillation (VF) than monophasic waveforms. If two phases are better than one, this naturally leads to the hypothesis that additional phases improve efficacy. This study tests the hypothesis by adding one additional phase. We examined the efficacy of 18 different triphasic waveforms simultaneo...

متن کامل

The influence of ventricular fibrillation duration on defibrillation efficacy using biphasic waveforms in humans.

OBJECTIVES The purpose of this study was to prospectively investigate the influence of ventricular fibrillation (VF) durations of 5, 10 and 20 s on the defibrillation threshold (DFT) during implantable cardioverter-defibrillator (ICD) implantation. BACKGROUND Although the DFT using monophasic waveforms has been shown to increase with VF duration in humans, the effect of VF duration on defibri...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Circulation

دوره 98 20  شماره 

صفحات  -