Thrombogenicity and early vascular healing response in metallic biodegradable polymer-based and fully bioabsorbable drug-eluting stents.
نویسندگان
چکیده
BACKGROUND Acute thrombogenicity and re-endothelialization represent clinically relevant end points pertaining to the safety of coronary stents, which have not been compared among biodegradable polymer-based drug-eluting metallic stents and fully bioabsorbable scaffolds to date. METHODS AND RESULTS We investigated comparative outcomes with respect to acute thrombogenicity and re-endothelialization among thin-strut biodegradable polymer metallic everolimus eluting stents (EES), thick-strut fully bioabsorbable EES, thick-strut biodegradable polymer metallic biolimus-eluting stents and control bare metal stents. An ex-vivo porcine arterio-venous shunt model was used to assess platelet aggregation, whereas a healthy rabbit model of iliofemoral stent implantation was used to assess re-endothelialization and inflammation. Confocal microscopy was used to detect fluorescently labeled antibody staining directed against CD61/CD42b for the identification of aggregated thrombocytes, CD14/PM-1, and RAM-11 for identification of neutrophils and monocytes/macrophages. Endothelial recovery was assessed by scanning electron microscopy, whereas CD31/PECAM-1 was used to confirm endothelial maturity. EES demonstrated significantly less acute thrombogenicity compared with bioabsorbable EES and biolimus-eluting stents. EES showed greater re-endothelialization at 28 days and reduced inflammatory cell adhesion of monocytes/macrophages at 14 days compared with bioabsorbable EES. Only bare metal stents showed complete re-endothelialization at 28 days. CONCLUSIONS These outcomes indicate differential trends in thrombogenicity and vascular healing among contemporary stents used in clinical practice and suggest a need for long-term adjunct antithrombotic pharmacotherapy for bioabsorbable EES.
منابع مشابه
Progress in treatment by percutaneous coronary intervention: the stent of the future.
First generation drug-eluting stents have considerably reduced in-stent restenosis and broadened the applications of percutaneous coronary interventions for the treatment of coronary artery disease. The polymer is an integral part of drug-eluting stents in that, it controls the release of an antiproliferative drug. The main safety concern of first generation drug-eluting stents with permanent p...
متن کاملRandomized comparison of biodegradable polymer sirolimus-eluting stents versus durable polymer everolimus-eluting stents for percutaneous coronary revascularization: rationale and design of the BIOSCIENCE trial.
BACKGROUND Biodegradable polymers for release of antiproliferative drugs from metallic drug-eluting stents aim to improve long-term vascular healing and efficacy. We designed a large scale clinical trial to compare a novel thin strut, cobalt-chromium drug-eluting stent with silicon carbide-coating releasing sirolimus from a biodegradable polymer (O-SES, Orsiro; Biotronik, Bülach, Switzerland) w...
متن کاملEvaluation of a novel slow-release paclitaxel-eluting stent with a bioabsorbable polymeric surface coating.
OBJECTIVES We sought to evaluate a new second-generation drug-eluting stent (DES), comprising a slow-release biodegradable polylactide coglycolide (PLGA) polymer and low-dose paclitaxel on a thin-strut cobalt chromium stent platform, in a clinically relevant animal model. BACKGROUND Our previous work demonstrated subacute vascular toxicity and necrosis triggering late excess neointima in pig ...
متن کاملImpact of stent surface on thrombogenicity and vascular healing: a comparative analysis of metallic and polymeric surfaces.
BACKGROUND Emerging drug-eluting stent technologies are evolving toward the elimination of polymeric component used as the method for modulating drug delivery. Although this technological approach seems to be biologically appealing, the impact of durable polymers and metallic stent surfaces on vascular healing remains unclear. In the present study, we aimed to compare the independent effect of ...
متن کاملCan Long-Term Results Following Balloon Angioplasty Be the “Crystal Ball” to Predict Outcome Following Bioresorbable Vascular Scaffolds?
I n this issue, Yamaji et al 1 report the very long-term (15-year) outcomes of patients treated successfully with balloon angioplasty (BA) compared with bare-metal stent implantation (BMS). Considering the major progresses made in interventional cardiology since the advent of BA, it would be reasonable to question what relevance these data have in the era of thirdand fourth-generation drug-elut...
متن کاملذخیره در منابع من