Growth detection failures by the nonradiometric Bactec MGIT 960 mycobacterial culture system.

نویسندگان

  • Jeremy A Peña
  • Mary Jane Ferraro
  • Colleen G Hoffman
  • John A Branda
چکیده

Mycobacterial growth in liquid culture can go undetected by automated, nonradiometric growth detection systems. In our laboratory, instrument-negative tubes from the Bactec MGIT 960 system are inspected visually for clumps suggestive of mycobacterial growth, which (if present) are examined by acid-fast smear analysis. A 3-year review demonstrated that ∼1% of instrument-negative MGIT cultures contained mycobacterial growth and that 10% of all cultures yielding mycobacteria were instrument negative. Isolates from instrument-negative MGIT cultures included both tuberculous and nontuberculous mycobacteria.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Unreliable detection of Mycobacterium xenopi by the nonradiometric Bactec MGIT 960 culture system.

From June 2006 to December 2007, 3,648 clinical specimens consecutively received for mycobacterial culture were investigated. Each processed sample was inoculated into Bactec MGIT 960 liquid medium and a Löwenstein-Jensen slant. Tubes that were flagged as positive by the instrument as well as those determined to be negative after 42 days of incubation were removed, visually inspected for growth...

متن کامل

Use of BACTEC MGIT 960 for recovery of mycobacteria from clinical specimens: multicenter study.

The BACTEC MGIT 960 instrument is a fully automated system that exploits the fluorescence of an oxygen sensor to detect growth of mycobacteria in culture. Its performance was compared to those of the radiometric BACTEC 460 instrument and egg-based Lowenstein-Jensen medium. An identical volume of sample was inoculated in different media, and incubation was carried out for 6 weeks with the automa...

متن کامل

Evaluation of the BACTEC MGIT 960 and the MB/BacT systems for recovery of mycobacteria from clinical specimens and for species identification by DNA AccuProbe.

A total of 120 mycobacterial isolates were recovered from 1,068 clinical specimens. Of these, 82.5% were in MGIT 960, 83.3% were in MB/BacT, 80% were in BACTEC 460, and 70% were on Löwenstein-Jensen medium. Mean times to detection of Mycobacterium tuberculosis (n = 96) were significantly shorter with MGIT 960 (12.6 days, P = 0.003) and BACTEC 460 (11.8 days, P < 0.001) than with MB/BacT (15.9 d...

متن کامل

Comparison of recoveries of mycobacterium tuberculosis using the automated BACTEC MGIT 960 system, the BACTEC 460 TB system, and Löwenstein-Jensen medium.

Using two different liquid media and one conventional solid medium, a total of 57 mycobacterial isolates (Mycobacterium tuberculosis, n = 55; nontuberculous mycobacteria, n = 2) were recovered from 377 clinical specimens. The rates of recovery of M. tuberculosis were 96. 4% with the BACTEC MGIT 960 liquid medium, 92.7% with BACTEC 12B liquid medium, and 81.8% with the Löwenstein-Jensen (LJ) med...

متن کامل

Factors Associated with Missed Detection of Mycobacterium tuberculosis by Automated BACTEC MGIT 960 System

Despite the demonstration of excellent performance, mycobacterial growth in BACTEC MGIT 960 can go undetected. The aim of this study was to investigate the prevalence of "false-negative" culture sample in Beijing and the potential factors associated with the detection failures by MGIT 960. Of the 577 sputum samples tested, 141 (24.4%) were culture-positive for mycobacteria, of which 133 (94.3%)...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of clinical microbiology

دوره 50 6  شماره 

صفحات  -

تاریخ انتشار 2012