Extraction of a well fixed but fractured ceramic acetabular liner
نویسندگان
چکیده
Ceramic fractures have been reported to occur in hip replacements, but the techniques used to extract a well fixed but fractured component are not commonly described. We present a case of ceramic liner fracture and validate a modification of a previously reported extraction technique that allowed us to save the acetabular cup and insert a polyethylene liner. With an increasing trend in ceramic bearing usage, it is likely that the number of ceramic liner fractures will increase and therefore knowledge of successful extraction techniques will be desirable.
منابع مشابه
A fractured and immovable ceramic liner in a screw-fixed acetabular shell during total hip arthroplasty: a case report.
We report a complication during total hip arthroplasty using an uncemented press-fit acetabular shell and ceramic liner for a woman with osteoarthritis. Fracture of the ceramic liner occurred during its insertion owing to non-concentric reduction. The main body of the ceramic liner became wedged in the acetabular shell and was resistant to removal. The shell itself was immovable owing to the pr...
متن کاملSurvival of ceramic bearings in total hip replacement after high-energy trauma and periprosthetic acetabular fracture.
Surgeons remain concerned that ceramic hip prostheses may fail catastrophically if either the head or the liner is fractured. We report two patients, each with a ceramic-on-ceramic total hip replacement who sustained high-energy trauma sufficient to cause a displaced periprosthetic acetabular fracture in whom the ceramic bearings survived intact. Simultaneous fixation of the acetabular fracture...
متن کاملAdverse local tissue reaction after 2 revision hip replacements for ceramic liner fracture
INTRODUCTION In younger patients, ceramic-on-ceramic (CoC) bearing surfaces are usually recommended for total hip replacement (THR) because of their low wear rate and longer expected functional life. Although technical advancements have reduced the risk of ceramic bearings fracture, this complication remains a major concern. CASE DESCRIPTION We present the case of a 56-year-old patient underg...
متن کاملLate liner disassociation of a Pinnacle system acetabular component.
This article describes a case of late locking mechanism failure and disassociation of a Pinnacle acetabular cup (DePuy, Warsaw, Indiana) and Marathon polyethylene liner (DePuy) 53 months after routine primary total hip arthroplasty in an active patient. Following an uncomplicated initial postoperative recovery, the patient felt a pop while kneeling for gardening activities. The patient had no p...
متن کاملCatastrophic ceramic failure in total hip arthroplasty: the role of microseparation
In May 2008, a left total hip arthroplasty was undertaken for the treatment of end-stage osteoarthritis in a 77-year-old man with a body mass index of 25 kg/m. A cementless total hip arthroplasty prosthesis was implanted through a transgluteal approach. A combination of a ceramic acetabular liner with a 36 mm ceramic femoral head was chosen. During the procedure, a low femoral neck cut resulted...