[Characteristic evaluation and clinical usefulness of dose area product meter at radiography].
نویسندگان
چکیده
Philips DigitalDiagnost, a digital radiographic system mounting flat panel detector (FPD), can display dose area data (DiDi dose) calculated by examination parameters. We evaluated its fundamental characteristics and compared the values of DiDi dose andactual measured data obtained by the area dose product meter (PD-4100L). Tendency of varied values of mAs, X-ray tube values and exposure area from both the area dose product meter and the DiDi dose were coincided. Further, in clinical images of chest PA 100 cases, chest lateral 50 cases and abdomen stand 25 cases, the determination coefficient was overly high as R(2)=0.99. Based on these results, it is clear that the DiDi dose can be treated the same as the area dose product meter. Under increasing of patient X-ray exposure dose is a concern in digital general radiography, this research indicates that maximum values of histogram obtained by DiDi dose contributes dose awareness for radiographer.
منابع مشابه
The Practice of Chest Radiography Using Different Digital Imaging Systems: Dose and Image Quality
Introduction: The study was undertaken to evaluate the practice of chest radiography using different digital imaging systems and its influence on dose and image quality. Materials and Methods: The study was carried out in two hospitals from March 2016 to June 2016. Sixty ambulatory patients aged 21 to 60 years who were able to cooperate without difficulty and weighed between 60 to 80 kg were ...
متن کاملA Comparison of Patients Absorption Doses with Bone Deformity Due to the EOS Imaging and Digital Radiology
Background: This study has aimed to measure the patient dose in entire spine radiography by EOS system in comparison with the digital radiography. Methods: EOS stereo-radiography was used for frontal and lateral view spine imaging in 41 patients in a prospective analytical study. A calibrated dose area product (DAP) meter was used for calibration of the DAP in EOS system. The accuracy and pr...
متن کاملA Phantom Study for the Optimization of Image Quality and Radiation Dose for Common Radiographic Examinations in Digital Radiography
Introduction: Phantom studies facilitate the implementation of radiation dose surveillance as a function of radiographic technical parameters for minimizing patient radiation dose. The evidence of such investigations can then be used to evaluate technical parameters used in the radiographic procedures to reduce radiation dose without compromising the image quality. Material and Methods: This ex...
متن کاملEvaluation of the Influence of Exposure Index on Image Quality and Radiation Dose
Introduction: The introduction of digital radiography has led to a significant problem in terms of dose creep. To address this problem, manufacturers have established a set of exposure indicators (EI) as a feedback mechanism to safeguard against overexposure. The EI is the measure of incident exposure to the detector that is directly proportional to the signal-to-noise...
متن کاملApplication of Dose-Area Product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology
Background: Application of Dose-area product (DAP) quantity has been increased in the clinical practice. DAP is relatively easy to measure. DAP has been shown to correlate well with the total energy imparted to the patient, which is related to the effective dose. Material and Methods: In this study, measurements of DAP were carried out in a sample of 421 adult patients who underwent convention...