Measuring containment of viable infectious cell sorting in high-velocity cell sorters.
نویسندگان
چکیده
BACKGROUND With the advent of high-speed sorters, aerosols are a considerable safety concern when sorting viable infectious materials. We describe a four-part safety procedure for validating the containment. METHODS This procedure includes aerosol containment, physical barriers, environmental controls, and personal protection. The Aerosol Management System (AMS) produces a negative pressure within the sort chamber, where aerosols are forced through a HEPA filter. Physical barriers include the manufacturer's standard plastic shield and panels. The flow cytometer was contained within a BSL-3 laboratory for maximum environmental control, and the operator was protected by a respiratory system. Containment was measured by using highly fluorescent Glo-Germ particles under the same conditions as the cell sort. RESULTS Escaping aerosols were vacuumed for 10 min onto a glass slide and examined. With the AMS active and the cytometer producing the maximum aerosols possible, Glo-Germ particles remained within the sort chamber. Measurements taken directly outside the door averaged fewer than one particle per slide, and those taken at 2 ft away and on top of the sorter were completely negative. CONCLUSIONS With this monitoring system in place, aerosols can be efficiently measured, thus reducing the risk to the operator while sorting viable infectious cells.
منابع مشابه
International Society for Analytical Cytology biosafety standard for sorting of unfixed cells.
BACKGROUND Cell sorting of viable biological specimens has become very prevalent in laboratories involved in basic and clinical research. As these samples can contain infectious agents, precautions to protect instrument operators and the environment from hazards arising from the use of sorters are paramount. To this end the International Society of Analytical Cytology (ISAC) took a lead in esta...
متن کاملStandard safety practices for sorting of unfixed cells.
Cell sorting of viable biological specimens has become widespread in laboratories involved in basic and clinical research. As these samples can contain infectious agents, precautions to protect instrument operators and the environment from hazards arising from the use of sorters are paramount. This unit presents a revised and updated version of the biosafety guidelines for sorting of unfixed ce...
متن کاملNovel rapid method for visualization of extent and location of aerosol contamination during high-speed sorting of potentially biohazardous samples.
BACKGROUND Containment of potentially biohazardous aerosols that result from high-speed sorting of human cells has been an increasingly important problem in analytical cytometry. The current method for assessing the efficiency of aerosol containment involves detection of aerosols containing sorted T4 bacteriophage on lawns of T4-susceptible Escherichia coli on plates that are placed in and arou...
متن کاملInternational Society for the Advancement of Cytometry cell sorter biosafety standards.
Flow cytometric cell sorting of biological specimens has become prevalent in basic and clinical research laboratories. These specimens may contain known or unknown infectious agents, necessitating precautions to protect instrument operators and the environment from biohazards arising from the use of sorters. To this end the International Society of Analytical Cytology (ISAC) was proactive in es...
متن کاملOptimizing Optical Flow Cytometry for Cell Volume-Based Sorting and Analysis
Cell size is a defining characteristic central to cell function and ultimately to tissue architecture. The ability to sort cell subpopulations of different sizes would facilitate investigation at genomic and proteomic levels of mechanisms by which cells attain and maintain their size. Currently available cell sorters, however, cannot directly measure cell volume electronically, and it would the...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Cytometry. Part A : the journal of the International Society for Analytical Cytology
دوره 52 2 شماره
صفحات -
تاریخ انتشار 2003