Single-Lead DDD System: A Comparative Evaluation of Unipolar, Bipolar and Overlapping Biphasic Stimulation

نویسندگان

  • H. - F. TSE
  • C. P. LAU
  • Z. LEUNG
چکیده

Single-lead DDD pacing using unipolar or bipolar stimulation has been limited by high atrial thresholds. Overlapping biphasic (OLBI) waveform stimulation via atrial floating ring electrodes may preferentially enhance atrial stimulation and avoid diaphragmatic pacing. Single-lead DDD pacing using the OLBI pacing configuration was studied in 12 patienta (6 male, 6 female, mean age 74±7 years) with complete heart block. At implantation, atrial rings (area 27 mm2, separation 10 mm) were positioned at radiological defined high, mid and low right atrium (RA). The P-wave amplitude, atrial and diaphragmatic pacing thresholds were determined at each position using unipolar, bipolar and OLBI stimulation configurations in randomized order. Although statistically insignificant, the P-wave amplitude tended to be lower in the low RA position. Independent of the stimulation modes, minimum atrial pacing threshold occured in the mid RA. In the mid RA position , the atrial pacing threshold was significantly lower with the OLBI mode compared to either unipolar or bipolar mode (3,9±2,2V vs. 6,7±3,5V vs. 6,9±3,5V, p < 0,05). Althrough the diaphragmatic thresholds were similar, OLBI pacing modes in mid RA and final location sigificantly improved the safety margin for preventing diaphragmatic pacing compared to unipolar mode. By the time of pre-discharge testing, one patient developed atrial fibrillation, the remaining 11 patients had satisfactory atrial capture and stable atrial pacing threshold (day 0: 2,6±1,1V vs. day 2: 3,2±1,3V, p = ns). However, diaphragmatic pacing occured in 4/11 (36%) of patients especially during upright postures (sitting and standing). Preliminary clinical results suggest that OLBI pacing via atrial floating ring electrodes can reduce atrial pacing treshold compared to unipolar and bipolar configurations. To optimize atrial pacing and sensing, the bipolar electrodes should be located at the mid RA position first, althrough the right RA may be alternative. Despite significant improvement of safety marginfor diaphragmatic pacing with OLBI pacing mode, diaphragmatic stimulation remains a clinical problem.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Single-Lead DDD Pacing with Two Ring Electrodes in the Right Atrium and One Ring in the Superior Vena Cava

Dual-chamber (DDD) pacing for maintenance of atrioventricular (AV) synchrony generally requires the implantation of two leads, one in the right atrium and one in the right ventricle. During the early 1990's, the efforts were made to achieve DDD pacing using only one lead [1-3]. The idea was to use the electrical field generated between the floating ring electrodes located in the right atrium to...

متن کامل

Asymmetric Rectangular Waveform in Stimulation with High Frequency Alternating Current Reduces the Threshold for Neural Conduction Block

Introduction Abnormal neural impulses in the nervous system may lead to various diseases and disabilities. High frequency alternating currents (HFAC) has been used to block the propagation of such impulses and improve the symptoms or disabilities. The technique is safe, reversible, and relatively selective, and its reliability, the optimum stimulation parameters, and elimination of the onset re...

متن کامل

A Prospective, Randomized Study

Background Recent development of a prototype singlelead unipolar transvenous defibrillator offers the possibility of device implantation with the ease of a permanent pacemaker. Lowering defibrillation energy requirements would allow for a further reduction in defibrillator generator size and enhance the feasibility of pacemaker-like placement. However, if achieving a lower defibrillation energy...

متن کامل

DIAGNOSTIC METHODS ELECTROPHYSIOLOGY Comparison of bipolar and unipolar programmed electrical stimulation for the initiation of ventricular arrhythmias: significance of anodal excitation during bipolar stimulation

To determine if anodal excitation during bipolar stimulation facilitates the initiation of sustained monomorphic ventricular tachycardia, nonsustained polymorphic ventricular tachycardia, or repetitive ventricular responses, both bipolar and cathodal unipolar programmed ventricular stimulation with one to three extrastimuli delivered during ventricular pacing at two rates from the right ventric...

متن کامل

Comparison of the efficacy of unipolar and bipolar electrode configuration during subthalamic deep brain stimulation.

Deep brain stimulation of the subthalamic nuclei (STN) is a well established treatment in advanced Parkinson's disease (PD). Based on the clinical efficacy and elicited side-effects, both unipolar and bipolar stimulation modes may be applied. Bipolar stimulation usually produces a more focused and therefore thinner area of tissue activated during stimulation than unipolar stimulation does. The ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2002