Heterophilic antibodies in human sera causing falsely increased results in the CA 125 immunofluorometric assay.
نویسندگان
چکیده
An in-house OC 125 monoclonal antibody-based "sandwich" immunofluorometric assay (IFMA) described previously (Clin Chem 1987; 33:2191-4) gave higher results for CA 125 in 37 of 123 serum samples than did a commercially available immunoradiometric assay (IRMA). Discordant results between the two assays became concordant when measurements of the samples were repeated with normal mouse serum (100 mL/L) included in the IFMA reagent. The murine immunoglobulins are thought to block the ability of the heterophilic antibodies in the human serum samples to cross-link the labeled antibody with the solid-phase antibody. Using an enzyme immunoassay, we demonstrated human anti-mouse antibodies (HAMA) in most of the discordant samples examined. We tested the ability of nonimmune sera from other animal species to lower the apparent CA 125 concentrations of the spurious samples and observed that rat, goat, and sheep serum were less effective than mouse serum. One serum sample was discovered to give a falsely increased CA 125 result with the IRMA, but this increase could be prevented by adding murine serum to the IRMA reagent. We conclude that falsely increased CA 125 results are best prevented by adding murine serum (or murine antibodies) to the assay buffer.
منابع مشابه
Interference by human anti-mouse antibodies in CA 125 assay after immunoscintigraphy: anti-idiotypic antibodies not neutralized by mouse IgG but removed by chromatography.
Falsely increased concentrations of the ovarian carcinoma-associated antigen, CA 125, were measured by a monoclonal antibody (MAb)-based double determinant immunoradiometric assay (IRMA) in patients who developed antibodies to mouse immunoglobulins (IgGs) after receiving injections of the same MAb as is used in the CA 125 IRMA. Addition of undiluted mouse serum or purified mouse IgG to the assa...
متن کاملHeteroantibody: Phantom of the Immunoassay
The heterophilic antibody, or heteroantibody, has been haunting immunoassays for many years. Interference by heteroantibodies in human sera can either increase or decrease the results of an immunoassay. For instance, by binding to a site other than the analyte-binding site, the heteroantibodies are capable of cross-linking the signal antibody with the capture antibody, thereby generating a fals...
متن کاملFalse elevation of serum CA-125 level caused by human anti-mouse antibodies.
Discordant results were observed for serum CA-125 (carbohydrate antigen-125) assays in a patient who was monitored for recurrence of ovarian cancer. Serum CA-125 levels in this patient were normal when measured in one laboratory, but >5-times the upper limit of normal (35 U/mL) when measured in another laboratory. Both laboratories used dual antibody heterogeneous immunoassays, but from differe...
متن کاملMisleading FT4 measurement: Assay-dependent antibody interference
Introduction: Commonly used free thyroxine (FT4) immunoassays can be falsely elevated due to interference causing misinterpreted thyroid function. We present two cases with high FT4 concentrations due to antibody interference. This study’s aim was to investigate the source of the FT4 immunoassay interference and possibility of its removal by two different techniques in order to correct the disc...
متن کاملEvidence for interaction of human anti-idiotypic antibodies with CA 125 determination in a patient after radioimmunodetection.
Very high concentrations of CA 125 have been found in some ovarian cancer patients after repeated radioimmunodetection with anti-CA 125 antibodies [OC125-F(ab')2]. In one patient we measured a CA 125 concentration of 135,000 kilo-arb. units/L, using an enzyme immunoassay involving OC125 antibodies. With an immunoradiometric assay involving use of two new anti-CA 125 antibodies (B43.13 and B27.1...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 36 6 شماره
صفحات -
تاریخ انتشار 1990