Third-generation FloTrac/Vigileo does not reliably track changes in cardiac output induced by norepinephrine in critically ill patients.

نویسندگان

  • X Monnet
  • N Anguel
  • M Jozwiak
  • C Richard
  • J-L Teboul
چکیده

BACKGROUND The ability of the third-generation FloTrac/Vigileo software to track changes in cardiac index (CI) induced by volume expansion and norepinephrine in critically ill patients is unknown. METHODS In subjects with circulatory failure, we administered volume expansion (20 subjects) and increased (20 subjects) or decreased (20 subjects) the dose of norepinephrine. We measured arterial pressure waveform-derived CI provided by the third-generation FloTrac/Vigileo device (CI(pw)) and transpulmonary thermodilution CI (CI(td)) before and after therapeutic interventions. RESULTS Considering the pairs of measurements performed before and after all therapeutic interventions (n=60), a bias between the absolute values of CI(pw) and CI(td) was 0.26 (0.94) litre min(-1) m(-2) and the percentage error was 54%. Changes in CI(pw) tracked changes in CI(td) induced by volume expansion with moderate accuracy [n=20, bias=-0.11 (0.54) litre min(-1) m(-2), r(2)=0.26, P=0.02]. When changes in CI(td) were induced by norepinephrine (n=40), a bias between CI(pw) and CI(td) was 0.01 (0.41) litre min(-1) m(-2) (r(2)=0.11, P=0.04). The concordance rates between changes in CI(pw) and CI(td) induced by volume expansion and norepinephrine were 73% and 60%, respectively. The bias between changes in CI(pw) and CI(td) significantly correlated with changes in total systemic vascular resistance (r(2)=0.41, P<0.0001). CONCLUSIONS The third-generation FloTrac/Vigileo device was moderately reliable for tracking changes in CI induced by volume expansion and poorly reliable for tracking changes in CI induced by norepinephrine.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of pulse contour analysis by Pulsioflex and Vigileo to measure and track changes of cardiac output in critically ill patients.

BACKGROUND We compared the new Pulsioflex and the Vigileo devices to measure cardiac index (CI) in critically ill patients. Both devices measure CI by pulse-contour analysis. The Pulsioflex device also allows an auto-calibration (not based on thermodilution). METHODS Patients were included if we administered fluids (20 patients), reduced (20 patients), or increased (20 patients) the dose of n...

متن کامل

Assessment of cardiac output changes using a modified FloTrac/Vigileo™ algorithm in cardiac surgery patients

INTRODUCTION The FloTrac/Vigileo (Edwards Lifesciences, Irvine, CA, USA) allows pulse pressure-derived cardiac output measurement without external calibration. Software modifications were performed in order to eliminate initially observed deficits. The aim of this study was to assess changes in cardiac output determined by the FloTrac/Vigileo system (FCO) with an initially released (FCOA) and a...

متن کامل

Performance of cardiac output measurement derived from arterial pressure waveform analysis in patients requiring high-dose vasopressor therapy.

BACKGROUND Arterial pressure waveform analysis of cardiac output (APCO) without external calibration (FloTrac/Vigileo™) is critically dependent upon computation of vascular tone that has necessitated several refinements of the underlying software algorithms. We hypothesized that changes in vascular tone induced by high-dose vasopressor therapy affect the accuracy of APCO measurements independen...

متن کامل

Unreliable Tracking Ability of the Third-Generation FloTrac/Vigileo™ System for Changes in Stroke Volume after Fluid Administration in Patients with High Systemic Vascular Resistance during Laparoscopic Surgery

BACKGROUND The FloTrac/Vigileo™ system does not thoroughly reflect variable arterial tones, due to a lack of external calibration. The ability of this system to measure stroke volume and track its changes after fluid administration has not been fully evaluated in patients with the high systemic vascular resistance that can develop during laparoscopic surgery. METHODS In 42 patients undergoing...

متن کامل

Fluid responsiveness prediction using Vigileo FloTrac measured cardiac output changes during passive leg raise test

BACKGROUND Passive leg raising (PLR) is a so called self-volume challenge used to test for fluid responsiveness. Changes in cardiac output (CO) or stroke volume (SV) measured during PLR are used to predict the need for subsequent fluid loading. This requires a device that can measure CO changes rapidly. The Vigileo™ monitor, using third-generation software, allows continuous CO monitoring. The ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • British journal of anaesthesia

دوره 108 4  شماره 

صفحات  -

تاریخ انتشار 2012