To examine the relationship between the defibrillation threshold and the strength of shocks that induce ventricular fibrillation during the vulnerable period, we determined the defibrillation threshold in 22 open-chest dogs using epicardial defibrillation electrodes with the cathode at the ventricular apex and the anode at the right atrium. We also determined whether there was an upper limit of shock strength that induces fibrillation in the vulnerable period by giving shocks of various energy through these same electrodes during the repolarization phase of paced rhythm. The above determinations were also made with the anode at the ventricular apex and the cathode at the right atrium in eight of the dogs and with the cathode at the ventricular apex and the anode at the left atrium in another eight of the dogs. In all dogs for all electrode configurations, there was an upper limit to the shock strength that induced ventricular fibrillation during the vulnerable period. Depending on the electrode combination, this upper limit of ventricular vulnerability either was not significantly different from or was slightly lower than the defibrillation threshold. The correlation coefficient between the two was highly significant for all three electrode configurations. These results support the hypothesis that successful defibrillation with epicardial electrodes requires a shock strength that reaches or exceeds the upper limit of ventricular vulnerability and that shocks slightly lower than the defibrillation threshold fail because they reinitiate ventricular fibrillation by stimulating portions of the myocardium during their vulnerable period.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Upper limit of vulnerability reliably predicts the defibrillation threshold in humans.

BACKGROUND The upper limit of vulnerability is the stimulus strength above which electrical stimulation cannot induce ventricular fibrillation even when the stimulus occurs during the vulnerable period of the cardiac cycle. The purpose of this study was to test the hypothesis that the upper limit of vulnerability can accurately predict the defibrillation threshold in patients undergoing implant...

متن کامل

Relation between upper limit of vulnerability and defibrillation threshold in humans.

BACKGROUND In the canine model, an upper limit of shock strength exists that can induce ventricular fibrillation during the vulnerable period of the cardiac cycle. This shock strength (the upper limit of vulnerability) closely correlates with the defibrillation threshold and supports the "upper limit of vulnerability" hypothesis of defibrillation. It is not known whether an upper limit of vulne...

متن کامل

Upper Limit of Vulnerabilit Reliably Predicts the Defibrillation Threshold in Humans

Background The upper limit of vulnerability is the stimulus strength above which electrical stimulation cannot induce ventricular fibrillation even when the stimulus occurs during the vulnerable period of the cardiac cycle. The purpose of this study was to test the hypothesis that the upper limit of vulnerability can accurately predict the defibrillation threshold in patients undergoing implant...

متن کامل

Optical mapping of ventricular defibrillation in isolated swine right ventricles: demonstration of a postshock isoelectric window after near-threshold defibrillation shocks.

BACKGROUND Investigators who studied ventricular defibrillation by use of optical mapping techniques failed to observe an initial defibrillation event (isoelectric window or quiescent period) shown by electrode mapping studies. This discrepancy has important implications for the mechanisms of defibrillation. The purpose of the present study was to demonstrate an optical equivalent of an isoelec...

متن کامل

Inductionless or limited shock testing is possible in most patients with implantable cardioverter- defibrillators/cardiac resynchronization therapy defibrillators: results of the multicenter ASSURE Study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduction Evaluation With Implantable Cardioverter-Defibrillator Implantations).

BACKGROUND Implantable cardioverter-defibrillators and cardiac resynchronization therapy defibrillators have relied on multiple ventricular fibrillation (VF) induction/defibrillation tests at implantation to ensure that the device can reliably sense, detect, and convert VF. The ASSURE Study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduc...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Circulation

دوره 73 5  شماره 

صفحات  -

تاریخ انتشار 1986