Successful demonstration of simultaneous PET/MR Imaging with a RF-penetrable PET insert
نویسندگان
چکیده
Stanford University, Palo Alto, CA, USA The integration of PET’s ability to visualize and quantify molecular signatures and MRI’s excellent anatomical information shows great promise to be a powerful tool for disease characterization. However, the high cost of simultaneous PET/MRI has limited its availability. To address this problem, we have developed an RF-penetrable PET insert that can be placed in the bore of any MRI system without requiring modifications to the latter. The PET insert consists of 16 PET detector modules within Faraday cages in a 32 cm inner diameter ring pattern with small gaps. These gaps, along with electrooptical signal transmission technology that allows the PET insert to electrically float relative to the MRI RF ground, enable RF fields to pass through the PET ring with some attenuation. In this study, we investigated the feasibility of the RF-penetrable PET insert placed inside a 3T MR system using the body coils for the RF transmit/receive coil. We analyzed the transmit and receive penetration of the RF fields by acquiring B1 maps and MR images. When the PET insert was placed inside the MRI bore, compared to the case with no PET insert, the mean B1 field amplitude (transmitted RF field) was attenuated by -3.47 dB. The SNR (received MR signal) was attenuated by similar factors of -3.4 dB and -3.9 dB with GRE and FSE sequences, respectively. In addition, we acquired simultaneous PET/MR images. The PET images acquired inside the MRI bore were equivalent to those acquired outside the MR system. We have shown that the RF-penetrable PET technology allows the RF field of an MR body coil to penetrate into the field-of-view, enabling acquisition of simultaneous PET/MR images via a PET insert, without modification to the MR system.
منابع مشابه
PET/MRI insert using digital SiPMs: Investigation of MR-compatibility
In this work, we present an initial MR-compatibility study performed with the world's first preclinical PET/MR insert based on fully digital silicon photo multipliers (dSiPM). The PET insert allows simultaneous data acquisition of both imaging modalities and thus enables the true potential of hybrid PET/MRI. Since the PET insert has the potential to interfere with all of the MRI's subsystems (s...
متن کاملSmall animal simultaneous PET/MRI
We developed a non-magnetic positron-emission tomography (PET) device based on the rat conscious animal PET that operates in a small-animal magnetic resonance imaging (MRI) scanner, thereby enabling us to carry out simultaneous PET/MRI studies. The PET detector comprises 12 detector blocks, each being a 4 × 8 array of lutetium oxyorthosilicate crystals (2.22 × 2.22 × 5 mm3) coupled to a matchin...
متن کاملMR-compatibility study of a preclinical digital PET/MRI insert
Department of Physics of Molecular Imaging Systems, Institute for Experimental Molecular Imaging, RWTH Aachen University, Aachen, Germany The combination of PET and MRI into a hybrid device is challenging since both systems might interfere with each other. Therefore, the study of the MR-compatibility of such a combined device is crucial to understand and solve potential interference phenomena. ...
متن کاملDevelopment of PET insert for simultaneous PET/MR imaging of human brain
Molecular Imaging Research & Education (MiRe) Laboratory, Department of Electronic Engineering, Sogang University, Seoul, Korea Recently, there has been great interest on the development of combined PET/MR, which is a useful tool for both functional and anatomic imaging. The purpose of this study was to develop a MR compatible PET insert for simultaneous PET and MR imaging of human brain and to...
متن کاملFPGA-based RF interference reduction techniques for simultaneous PET–MRI
The combination of positron emission tomography (PET) and magnetic resonance imaging (MRI) as a multi-modal imaging technique is considered very promising and powerful with regard to in vivo disease progression examination, therapy response monitoring and drug development. However, PET-MRI system design enabling simultaneous operation with unaffected intrinsic performance of both modalities is ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 2 شماره
صفحات -
تاریخ انتشار 2015