T2-weighted MR Imaging of the Pelvis at 3.0 Tesla: Optimum routine Fast Spin-echo Sequences
نویسندگان
چکیده
INTRODUCTION: MR scanners at 3.0 T for the body have been introduced in clinical settings recently. Because of the differences in T1 and T2 relaxation times between 3.0T and 1.5 T, optimal parameters for MR imaging should be different between those [1]. The purpose of the study was to establish parameters for routine clinical T2-weighted fast spin-echo (FSE) MR sequences for the pelvis at 3.0 T. METHODS: T2-weighted MR images obtained with FSE sequences with three sets of imaging parameters in 25 patients (15 women and 10 men) were analyzed. The mean age of women was 51 years (age range, 28-68 years), and the mean age of men was 71 years (61-82 years). All patients underwent imaging with a whole-body 3.0-T MR scanner (Signa Excite 3.0T HD, GE Healthcare) equipped with an 8 channel coils. The three sets of parameters were as follows: (a) TR(ms)/TE(ms)/ETL=8000/80/20, (b) 6000/80/16, and (c) 4000/80/12. These sets were selected in consideration of various T2 contrast without significant differences in imaging times. A matrix of 512 x 256 512 x 320 pixels was used, with a field of view of 24 cm, section thickness of 4-6 mm, and NEX of 2. Imaging times were approximately 3-5 min. Image blurring was evaluated visually by a radiologist with a four-point scale (1=poor, 2=fair, 3=good, 4=excellent). Contrast-to-noise ratios between (a) the myometrium and the junctional zone of the uterus, and (b) the endometrium and the skeletal muscle were calculated for each woman, and that between (c) the skeletal muscle and the peripheral zone of the prostate was calculated for each man. RESULTS: There were no significant differences in image blurring among three sequences (median=3 for all sequences, P>0.05). Contrast-to-noise ratios (mean ± SD) between the myometrium and the junctional zone were 4.2 ± 3.6 (TR/TE/ETL=8000/80/20), 3.8 ± 2.8 (6000/80/16), and 3.4 ± 2.2 (4000/80/12) (Fig 1). There were no significant differences among the three sequences (P>0.05, Tukeys test). FSE images with TR of 8000 (19.2 ± 9.1, P=0.001) and with TR of 6000 (19.4 ± 8.4, P=0.001) showed higher contrast-to-noise ratios between the endometrium and the skeletal muscle than images with TR of 4000 (16.0 ± 7.1), though there was no significant differences between images with TR of 8000 and 6000 (P=0.98) (Fig 2). FSE images with TR of 8000 (18.0 ± 5.0) showed higher contrast-to-noise ratio between the skeletal muscle and the peripheral zone of the prostate than images with TR of 6000 (16.4 ± 4.1, P=0.04) and with TR of 4000 (14.0 ± 3.3, P<0.001) (Fig 3). Also, the difference between the images with TR of 6000 and those with TR of 4000 was statistically significant (P=0.003). DISCUSSION: T2-weighted MR imaging plays an important role in diagnosing malignant tumors in the uterus or the prostate. Thus, obtaining good T2 contrast is crucial for practical 3.0T MR imaging of the pelvis. Although optimal parameters for MR imaging might be different between 3.0T and 1.5T, many radiologists use the same TR and TE at 3.0T as at 1.5T [2]. The MR signal intensity (S) for spin-echo sequence is generally given by the following equation:
منابع مشابه
Comparison of Three-Dimensional Double-Echo Steady-State Sequence with Routine Two-Dimensional Sequence in the Depiction of Knee Cartilage
Introduction: There are some routine two-dimensional sequences, including short tau inversion recovery (STIR), T2-weighted fast-spin echo (T2W-FSE), and proton-density fast spin-echo for diagnosing osteoarthritis and lesions of the knee cartilage. However, these sequences have some disadvantages, such as long scan time, inadequate spatial resolution, and suboptimal tis...
متن کاملComparison the Accuracy of Fetal Brain Extraction from T2-Half-Fourier Acquisition Single-Shot Turbo Spin-Echo (HASTE) MR Image with T2-True Fast Imaging with Steady State Free Precession (TRUFI) MR Image by Level Set Algorithm
Background Access to appropriate images of fetal brain can greatly assist to diagnose of probable abnormalities. The aim of this study was to compare the suitability of T2-True Fast Imaging with Steady State Free Precession (T2-TRUFI), and T2-Half-Fourier Acquisition Single-Shot Turbo Spin-Echo (T2- HASTE( magnetic resonance imaging (MRI) to extract the fetal brain using the level set algorithm...
متن کاملA comparison of MR sequences for lesions of the parotid gland.
PURPOSE To compare six MR sequences (plain and gadolinium-enhanced fat suppressed T1-weighted spin echo, T2-weighted standard spin echo, fat-suppressed and non-fat-suppressed T2-weighted fast spin echo, and inversion-recovery T2-weighted fast spin echo) in their ability to detect, delineate, and characterize lesions of the parotid gland. METHODS Fifty-eight parotid gland lesions imaged on 47 ...
متن کاملThe role of classic spin echo and FLAIR sequences for the evaluation of myelination in MR imaging.
PURPOSE The aim of this study was to assess the features of the normal brain development in terms of myelination in infants and young children on fluid-attenuated inversion recovery (FLAIR) magnetic resonance (MR) imaging, and to determine if FLAIR imaging is superior to spin echo MR sequences. MATERIALS AND METHODS T1-weighted (T1W) fast spin echo T2-weighted (FSE T2W), and FLAIR images were...
متن کاملDetection of focal hepatic lesions with MR imaging: prospective comparison of T2-weighted fast spin-echo with and without fat suppression, T2-weighted breath-hold fast spin-echo, and gadolinium chelate-enhanced 3D gradient-recalled imaging.
OBJECTIVE The purpose of this study was to compare breath-hold three-dimensional (3D) rapid gradient-echo (GRE) MR imaging obtained before and after gadolinium chelate injection with T2-weighted fast spin-echo and T2-weighted breath-hold fast spin-echo (BHFSE) MR imaging in the detection of focal hepatic masses. SUBJECTS AND METHODS Fifty-three patients with 108 focal hepatic masses had, pros...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2007