The performance of BD FACSPresto™ for CD4 T-cell count, CD4% and hemoglobin concentration test in Ethiopia
نویسندگان
چکیده
INTRODUCTION In Ethiopia, CD4+ T-cell counting is still required for all patients at baseline before antiretroviral therapy (ART) and to determine eligibility and follow-up of opportunistic infection prophylaxis. However, access to CD4+ T cell count in rural health facilities remains a major challenge in Ethiopia like other resource-limited settings. METHODOLOGY Both capillary and venous blood was drawn from each of 325 study participant recruited in Addis Ababa and surroundings. The CD4+ T-cell count, CD4%, and hemoglobin (Hgb) were tested at one of the four study health facilities using capillary blood and BD FACSPresto™ device. These tests were also done at the national HIV reference laboratory, using venous blood with BD FACSCalibur™, Sysmex XT-1800i™, and BD FACSPresto™. RESULTS BD FACSPresto™ had an absolute mean bias of -13.3 cells/ul (-2.99%) and 28.3 cells/μl (6.4%) using venous and capillary blood, respectively, compared with BD FACSCalibur™. The absolute CD4 assay on the BD FACSPresto™ had a regression coefficient (R2) of 0.87 and 0.92 using capillary blood and venous blood samples, respectively, compared with BD FACSCalibur™. The percentage similarity of the BD FACSPresto™ using capillary and venous blood was 105.2% and 99.3%, respectively. The sensitivity of the FACSPresto™ using threshold of 500 cells/μl for ART eligibility using capillary and venous blood was 87.9 and 94.3%, while the specificity was 91.4 and 83.8%, respectively. Furthermore, the BD FACSPresto™ had an absolute mean bias of -0.2 dl/μl (0.0%) (95% LOA: -1.7, 1.3) and -0.59 dl/μl (0.1%) (95% LOA: -1.49, 0.31) for Hgb using capillary and venous blood compared with the Sysmex XT-1800i™, respectively. CONCLUSION Our results showed acceptable agreement between the BD FACSPresto™ and BD FACSCalibur™ for CD4+ T-cell counting and CD4%; and between the BD FACSPresto™ and Sysmex XT-1800i™for measuring Hgb concentration.
منابع مشابه
CD4 Lymphocyte Enumeration and Hemoglobin Assessment Aid for Priority Decisions: A Multisite Evaluation of the BD FACSPresto™ System
Background The BD FACSPresto™ system uses capillary and venous blood to measure CD4 absolute counts (CD4), %CD4 in lymphocytes, and hemoglobin (Hb) in approximately 25 minutes. CD4 cell count is used with portable CD4 counters in resource-limited settings to manage HIV/AIDS patients. A method comparison was performed using capillary and venous samples from seven clinical laboratories in five co...
متن کاملPerformance evaluation of BD FACSPresto™ point of care CD4 analyzer to enumerate CD4 counts for monitoring HIV infected individuals in Nigeria
BACKGROUND Despite the upsurge in support and intervention of donor agencies in HIV care and treatment programing in Sub-Sahara African, antiretroviral (ART) programs are still confronted with access and coverage challenges which influence enrolment of new patients. This study investigated the validity of point of care BD FACSPresto™ CD4 analyzer for CD4+ cell count, overall agreement, correlat...
متن کاملClinical Evaluation of the BD FACSPresto™ Near-Patient CD4 Counter in Kenya
BACKGROUND The BD FACSPresto™ Near-Patient CD4 Counter was developed to expand HIV/AIDS management in resource-limited settings. It measures absolute CD4 counts (AbsCD4), percent CD4 (%CD4), and hemoglobin (Hb) from a single drop of capillary or venous blood in approximately 23 minutes, with throughput of 10 samples per hour. We assessed the performance of the BD FACSPresto system, evaluating a...
متن کاملPerformance of FACSPresto Point-of-Care Instrument for CD4-T Cell Enumeration in Human Immunodeficiency Virus (HIV)-Infected Patients Attending Care and Treatment Clinics in Belgium and Tanzania
BACKGROUND CD4 T-cell counts are widely used to assess treatment eligibility and to follow-up HIV-infected patients. The World Health Organization prequalification of in vitro diagnostics program conducted a performance evaluation of the FACSPresto (BD Biosciences), a new point-of-care instrument to measure absolute CD4-T cell (CD4) counts and percentages in venous and capillary blood samples f...
متن کاملتعیین زیر جمعیت های لنفوسیت T و B و NK بکمک آنتی بادی های مونوکلونال در بیماران مبتلا به بیماری بهجت در ایران
Behcet disease (BD) is a systemic inflammatory disease of the unknown etiology. There is however, some evidence to suggest that immunological abnormalities are important in its pathogenesis, furthermore several T-cell abnormalities which may be quite relevant to autoimmune origin of the disease have been described. We report here our study of T-cell subsets, B and NK cells in 68 patients with B...