How certain drugs interfere negatively with the Jaffé reaction for creatinine.
نویسندگان
چکیده
Phenacemide and cephalothin may interfere either positively or negatively with the determination of creatinine by the Jaffé reaction, depending on the analytical system. The drug-picrate chromophores maximally absorb within the first minute of reaction (21 s for phenacemide, 45 s for cephalothin), after which the absorbances decrease. Thus, these drugs negatively interfere in analytical systems involving kinetic measurements when the time interval of measurement includes the period of decreasing absorbance.
منابع مشابه
Interferences in current methods for measurements of creatinine.
We measured the interference of carbonyl compounds, drugs, and other substances in human serum on the determination of creatinine by the two-point, fixed-time kinetic modification of the Jaffé reaction as well as by four enzymatic methods. We added known concentrations of the interfering substances to a solution of creatinine in water. For bilirubin, we used both pooled normal sera with added b...
متن کاملMechanism of cefoxitin and cephalothin interference with the Jaffé method for creatinine.
Some cepha antibiotics interfere positively in the Jaffé method for determining creatinine. The extent of this interference differs with the modification used, and its mechanism is unknown. We studied the spectrophotometric and thermodynamic properties of the reaction of picrate with creatinine and two cepha antibiotics, cefoxitin and cephalothin. The absorbance spectra of the chromophore produ...
متن کاملInfluence of endogenous and exogenous Jaffé-reaction-positive pseudo-creatinine chromogens in determination of creatinine in blood of normal and cephaloridine-administered rats.
To investigate a magnitude of analytic errors brought by certain interfering substances to the chemical method using Jaffé-reaction for determination of creatinine, two sorts of values determined by the chemical method and by the HPLC method were compared between each other, in blood samples obtained from control and cephaloridine (CER)-administered rats. The rats that received an intravenous i...
متن کاملNitromethane interferes in assay of creatinine by the Jaffé reaction.
A patient who ingested a methanol/nitromethane mixture (model airplane fuel) showed an apparent plasma creatinine concentration of 8.0 mmol/L by the Jaffé reaction when the actual creatinine concentration, as measured by a specific enzymatic method, was 0.09 mmol/L. This effect was due to nitromethane in the plasma. Interference by nitromethane with the determination of creatinine by reaction w...
متن کاملAn automated dry-slide enzymatic method evaluated for measurement of creatinine in serum.
We evaluated an automated dry-slide enzymatic method involving creatinine iminohydrolase for measurement of creatinine in serum with the Kodak Ektachem analyzer. The means (and SD) for three commercially available quality-control sera, analyzed during eight weeks, were 9.7 (1.0), 16.6 (0.9), and 61 (2.1) mg/L. The regression equation for 105 samples measured with the Technicon SMAC (x) and Ekta...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 31 2 شماره
صفحات -
تاریخ انتشار 1985