Convolution-Based Truncation Correction for C-Arm CT Using Scattered Radiation
نویسندگان
چکیده
Patient dose reduction in C-arm computed tomography by volume-of-interest (VOI) imaging is becoming an interesting topic for many clinical applications. One limitation of VOI imaging that remains is the truncation artifact in the reconstructed 3-D volume. This artifact can either be a cupping effect towards the boundaries of the field-of-view (FOV) or an offset in the Hounsfield values of the reconstructed voxels. A new method for the correction of truncation artifacts in a collimated scan is introduced in this work. Scattered radiation still reaches the detector and is detected outside of the FOV, even if axial or lateral collimation is used. By reading out the complete detector area, we can use the scatter signal to estimate the truncated parts of the object: The scattered radiation outside the FOV is modeled as a convolution with a scatter kernel. This new approach is called Scatter Correction. The reconstruction results using Scatter Convolution are at least as good or better than the results with a state-of-the-art method. Our results show that the use of scattered radiation outside the FOV improves image quality by 1.8%.
منابع مشابه
Calculation of the Scattered Radiation Profile in 64 Slice CT Scanners Using Experimental Measurement
Introduction: One of the most important parameters in x-ray CT imaging is the noise induced by detected scattered radiation. The detected scattered radiation is completely dependent on the scanner geometry as well as size, shape and material of the scanned object. The magnitude and spatial distribution of the scattered radiation in x-ray CT should be quantified for development of robust scatter...
متن کاملI. INTRODUCTION Advances in model-based iterative reconstruction (IR) methods for x-ray CT and cone-beam CT (CBCT) imaging
C-arm cone-beam CT offers great potential in image-guided interventions, but conventional analytic reconstruction methods are associated with limited image quality, particularly for soft-tissue imaging. While model-based iterative reconstruction (IR) methods improve image quality and/or reduce radiation dose, long reconstruction time limits utility in clinical workflow. Additionally, in contras...
متن کاملComparison of state-of-the-art atlas-based bone segmentation approaches from brain MR images for MR-only radiation planning and PET/MR attenuation correction
Introduction: Magnetic Resonance (MR) imaging has emerged as a valuable tool in radiation treatment (RT) planning as well as Positron Emission Tomography (PET) imaging owing to its superior soft-tissue contrast. Due to the fact that there is no direct transformation from voxel intensity in MR images into electron density, itchr('39')s crucial to generate a pseudo-CT (Computed Tomography) image ...
متن کاملThe Added Value of Volume-of-Interest C-Arm CT Imaging during Endovascular Treatment of Intracranial Aneurysms.
BACKGROUND AND PURPOSE Successful endovascular treatment of intracranial aneurysms requires understanding the exact relationship of implanted devices to the aneurysm, parent artery, and other branch vessels during the treatment. Intraprocedural C-arm CT imaging has been shown to provide such information. However, its repeated use is limited due to increasing radiation exposure to the patient. T...
متن کاملEnhancement of mobile C-arm cone-beam reconstruction using prior anatomical models.
We demonstrate an improvement to cone-beam tomographic imaging by using a prior anatomical model. A protocol for scanning and reconstruction has been designed and implemented for a conventional mobile C-arm: a 9 inch image-intensifier OEC-9600. Due to the narrow field of view (FOV), the reconstructed image contains strong truncation artifacts. We propose to improve the reconstructed images by f...