Influence of dynamic leaks in volume-targeted pressure support noninvasive ventilation: a bench study.
نویسندگان
چکیده
INTRODUCTION The effect of leaks on volume-targeted pressure support noninvasive ventilation mode has only been tested with continuous simulated leaks. The objective of the study was to assess the influence of random leaks occurring either during inspiration or expiration. METHODS Analysis of the volume-targeted pressure support mode in 6 commercial ventilators with single-limb circuits and intentional leak in a bench study (restrictive model). Unintentional leaks were introduced through a mechanical system during inspiration (threshold valve with 2 levels of leaks) or during expiration (active valve). Results of delivered tidal volume (VT) and pressure support were externally recorded. A pre-set VT of 550 mL was programmed, with a wide range of pressure support values. RESULTS All the ventilators showed a deviation of delivered versus programmed VT below 10% in the period without unintentional leaks. In the model with unintentional inspiratory leaks, a progressive drop in delivered VT and pressure support was observed for all ventilators. The reduction in the delivered VT for the highest inspiratory leak ranged between 21 and 40%, corresponding to a decrease in pressure support between 3.09 and 10.15 cm H2O after 5 min. Conversely, in the expiratory model, increases in delivered VT and pressure support were observed, ranging between 16 and 33% and between 2.7 and 6.5 cm H2O, respectively. CONCLUSIONS The introduction of random leaks influences the performance of commercial ventilators with single-limb circuits and intentional leak. The decrease in delivered VT with inspiratory leaks reaches a magnitude that may have clinically important impacts.
منابع مشابه
Performance of ventilators for noninvasive positive-pressure ventilation in children.
The aim of the present study was to evaluate the performance characteristics of all the ventilators proposed for home noninvasive positive-pressure ventilation in children in France. The ventilators (one volume-targeted, 12 pressure-targeted and four dual) were evaluated on a bench which simulated six different paediatric ventilatory patterns. For each ventilator, the quality of the inspiratory...
متن کاملNeonatal and adult ICU ventilators to provide ventilation in neonates, infants, and children: a bench model study.
BACKGROUND Using a bench test model, we investigated the hypothesis that neonatal and/or adult ventilators equipped with neonatal/pediatric modes currently do not reliably administer pressure support (PS) in neonatal or pediatric patient groups in either the absence or presence of air leaks. METHODS PS was evaluated in 4 neonatal and 6 adult ventilators using a bench model to evaluate trigger...
متن کاملPerformance of ICU ventilators during noninvasive ventilation with large leaks in a total face mask: a bench study* **
OBJECTIVE Discomfort and noncompliance with noninvasive ventilation (NIV) interfaces are obstacles to NIV success. Total face masks (TFMs) are considered to be a very comfortable NIV interface. However, due to their large internal volume and consequent increased CO2 rebreathing, their orifices allow proximal leaks to enhance CO2 elimination. The ventilators used in the ICU might not adequately ...
متن کاملHarms of unintentional leaks during volume targeted pressure support ventilation.
BACKGROUND Volume targeted pressure support ventilation (VT-PSV) is a hybrid mode increasingly used to maintain a minimal tidal volume (VT) by automatically adjusting the level of inspiratory pressure. The objective of the study was to determine the ability of home ventilators to maintain the preset minimal VT during unintentional leaks in a VT-PSV mode. METHODS Seven ventilators were tested ...
متن کاملMonitoring of noninvasive ventilation by built-in software of home bilevel ventilators: a bench study.
BACKGROUND Current bilevel positive-pressure ventilators for home noninvasive ventilation (NIV) provide physicians with software that records items important for patient monitoring, such as compliance, tidal volume (Vt), and leaks. However, to our knowledge, the validity of this information has not yet been independently assessed. METHODS Testing was done for seven home ventilators on a bench...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Respiratory care
دوره 60 2 شماره
صفحات -
تاریخ انتشار 2015