Biomechanical comparison of dynamic hip screw, proximal femoral nail, cannulated screw, and monoaxial external fixation in the treatment of basicervical femoral neck fractures.
نویسندگان
چکیده
PURPOSE OF THE STUDY The objective of this study was to establish relative fixation strengths of proximal femoral nail (PFN), dynamic hip screw (DHS), monolateral external fixator (EF), and cannulated screw (CS) in basicervical hip fracture model. MATERIAL AND METHODS The study involved four groups of implanted composite proximal femoral synthetic bones of eight specimens per group; nailing with PFN, DHS, fixation with three cannulated screws, and EF. 70˚ osteotomy was performed to simulate a Pauwels Type 3 basicervical fracture. Minimum preload of 100 N was applied before loading to failure. The constructs were subjected to cyclic loading with 16˚ to midline from 100 N to 1,000 N for 10,000 cycles at 3Hz. Axial loading was applied at 10 mm/min until failure. Failure load, failure mode, and displacement were documented. RESULTS Mean failure load was 2182.5 ± 377.9 N in PFN group, 2008.75 ± 278.4 N in DHS group, 1941.25 ± 171.6 N in EF group, and 1551.6 ± 236.2 N in CS group. Average displacement was 15.6 ± 4.5 mm, 15.5 ± 6.7 mm, 11.7 ± 1.9 mm, and 15 ± 1.7 mm, respectively. No significant difference was noted among groups for fixation strength except CS group. All CS constructs failed during cyclic loading. CONCLUSION Our findings suggest that PFN, DHS and EF achieved higher fixation strengths than CS in basicervical fracture. PFN has higher failure loads and possesses biomechanical benefits for fixation of unstable basicervical fractures compared with DHS and EF. Key words: basicervical fracture, internal fixation, biomechanics.
منابع مشابه
Comparing Therapeutic Outcomes of Dynamic Hip Screw and Multiple Cannulated Screws as Internal Fixation in Femoral Neck Fractures
Background and purpose: Femoral neck fracture is associated with a high rate of side effects, especially in young people, who are usually affected by severe energy shocks. Results after injury depend on some factors, including displacement, fragility, bleeding rate, proper placement, and stable and strong fixation. The aim of this study was to compare dynamic hip screw (DHS) or multiple cannula...
متن کاملType II Intertrochanteric Fractures: Proximal Femoral Nailing (PFN) Versus Dynamic Hip Screw(DHS)
Background: Intertrochanteric fracture is one of the most common fractures of the hip especially in the elderly with osteoporotic bones, usually due to low-energy trauma like simple falls. Dynamic Hip Screw (DHS) is still considered the gold standard for treating intertrochanteric fractures by many. Not many studies compare the DHS with Proximal femoral nail (PFN), in Type II intertrochanteri...
متن کاملFinite Element Analysis for the Treatment of Proximal Femoral Fracture
Dynamic hip screw and gamma nail have been widely used to treat the patients with proximal femoral fractures, but clinical failures of those implants are still to be found. This study developed three-dimensional finite element models to investigate the biomechanical performances of the implants. Two kinds of commercially available implants (dynamic hip screw and gamma nail) and one newly design...
متن کاملShort-term results of surgical treatment with cephalomedullary nails for basicervical proximal femoral fractures.
The treatment of basicervical femoral fractures remains controversial. The aim of this study was to examine the efficacy of intramedullary nail use in the surgical treatment of basicervical fractures. In total, 28 patients with basicervical fractures treated with proximal femoral nails were examined retrospectively. Fracture healing was observed in all patients, who were followed at least for 6...
متن کاملProximal Femoral Nail - A Minimally Invasive Method for Stabilization of Pertrochanteric and Subtrochanteric Femoral Fractures
Fractures of the proximal femur are a big challenge in traumatology. Rapid strides in implant and instrumentations in quest of ideal fixation of pertrochanteric and subtrochanteric femoral fractures have made various options available like Fixed-angled nail plate, Screw plates, Angled blade plate, Dynamic Hip Screw etc. The present study aims to study the role of standard PFN as a minimal invas...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Acta chirurgiae orthopaedicae et traumatologiae Cechoslovaca
دوره 82 2 شماره
صفحات -
تاریخ انتشار 2015