Comparison of liver ADC measurements using breath-hold, free breath-hold and respiratory gating echoplanar diffusion-weighted imaging sequences using parallel imaging technique with different acceleration factors
نویسندگان
چکیده
Introduction: Echoplanar diffusion-weighted images (EP-DWI) with breath holding (BH) have been widely applied to evaluate the apparent diffusion coefficient (ADC) of liver. Clinically, the spatial coverage of DWI is usually limited by the duration (around 20 seconds) of breath hold. Free breathing (FB) and respiratory gating (RG) methods allow larger spatial coverage of liver in one scan. Comparison of liver ADC measurements in BH, FB, and RG EP-DWI at two folds of acceleration has been rarely investigated recently [1~3]. The aim of this preliminary study was to verify the measurement differences of liver ADC among clinically available BH, FB and RG methods with different acceleration factors. Materials and Methods: Seventeen volunteers without any hepatic lesion (11 men and 6 women; 43.12 ± 16.44 years) were enrolled. All liver scans were performed on a 1.5T MR scanner (GE Healthcare, Signa HDx, US) using a 12-channel phase-array body coil. Scanning parameters including resolution (128x128), slice number (8), slice thickness (8 mm), gap (1 mm), b values (0 sec/mm and 600 sec/mm) and NEX (3) were consistent in all diffusion protocols. Four methods including BH (TR/TE/scan time = 1650/60.1/20), FB 1650 (1650/60.1/20), FB 4000 (4000/60.1/48) and RG (4000~7000/60.1/50~110) with acceleration factors of 1 and 2 were applied. In each volunteer, five circular regions-of-interest (ROIs) were manually placed in the peripheral liver parenchyma to avoid partial volume effect of blood vessels and bile ducts (Fig. 1). A total of 375 pixels, 75 pixels in each ROI, were used for analysis of ADC value. Statistical analyses were performed by using SPSS 13.0 (SPSS, Chicago, Ill). Normality of the perfusion parameters was examined using Q-Q plots and Kolmogorov-Smirnov tests. Student t test was used for group comparisons of ADC. A P value of less than 0.05 was considered as statistically significant. Results: All EP-DWI sequences showed satisfactory imaging quality on both DWI and ADC maps without perceptible distortion. Ts parenchymal ADC values, however, showed significant differences among different pulse sequences with details as summarized in Fig. 2. Discussion & Conclusion: The liver ADC values acquired by RG method are significantly higher than that acquired by BH method no matter with acceleration or not. Our results are in consistent with Kwee’s and Taouli’s studies [1, 2]. Kandpal’s study, however, shows no difference of liver ADC values acquired by RG and BH methods [3]. Such discrepancy might be attributed to the different repetition times. While choosing the same repetition time, we find that the ADC values acquired by FB 1650 method do not differ from that acquired by BH method no matter with acceleration or not. The ADC values of liver parenchyma measured by accelerated EP-DWI are significantly higher than that measured by non-accelerated EP-DWI in BH, FB 4000 and RG methods. Our findings are in consistent with a recent research in parotid glands [4]. In conclusion, our study demonstrates significant quantification differences of liver ADC values among three different clinically available pulse sequences. Unlike FB 4000 and RG, FB 1650 method allows ADC measurement similar to BH method and therefore might serve as a substitute for BH method for those who have difficulty in breathing hold. Reference: [1] Kwee TC, et al. JMRI 2008, 28: 1141-1148. [2] Taouli B, et al. JMRI 2009, 30: 561568. [3] Kandpal H, et al. AJR 2009, 192: 915-922. [4] Juan CJ, et al. Radiology 2009, 253: 144-152. (a) (b) (c)
منابع مشابه
Handout - Extracranial Diffusion - weighted Imaging Page 1 of 4 Diffuse liver and kidney disease : value of Diffusion - weighted imaging ( DWI )
Diffusion acquisition parameters for imaging the upper abdomen Breath-hold or free breathing or respiratory triggered (using a navigator echo) SS EPI DWI sequence can be acquired before contrast injection, with the suggested following parameters: axial or coronal acquisition, fat suppression, tridirectional gradients using the following b-values: 0 (used as reference), 50 (black-blood sequence-...
متن کاملEvaluation of Free Breathing Versus Breath Hold Diffusion Weighted Imaging in Terms Apparent Diffusion Coefficient (ADC) and Signal-to-Noise Ratio (SNR) Values for Solid Abdominal Organs
BACKGROUND Our aim was to compare the apparent diffusion coefficient (ADC) values of normal abdominal parenchymal organs and signal-to-noise ratio (SNR) measurements in the same patients with breath hold (BH) and free breathing (FB) diffusion weighted imaging (DWI). MATERIAL/METHODS Forty-eight patients underwent both BH and FB DWI. Spherical region of interest (ROI) was placed on the right h...
متن کاملNon-breath-hold high b-value diffusion-weighted MRI with parallel imaging technique: apparent diffusion coefficient determination in normal abdominal organs.
PURPOSE To detect apparent diffusion coefficient (ADC) values in normal abdominal organs using non-breath-hold high b-value diffusion-weighted magnetic resonance imaging (DW-MRI) with a parallel imaging technique. MATERIALS AND METHODS A total of 50 patients with normal abdominal MRI findings were retrospectively enrolled. DW-MRI was performed with b-factors of 0, 500, and 1000 s/mm(2). Mean ...
متن کاملA comparison of a T1 weighted 3D gradient-echo sequence with three different parallel imaging reduction factors, breath hold and free breathing, using a 32 channel coil at 1.5 T. A preliminary study.
PURPOSE To investigate whether increasing temporal resolution with higher parallel imaging (PI) reduction factors (RF) in both breath-hold and free breathing approaches, using a non-contrast T1-weighted 3D gradient echo (GRE) sequence and a 32-channel phased array coil, permits diagnostic image quality, with potential application in patients unable to cooperate with breath-hold requirements. ...
متن کاملHighly accelerated T1-weighted abdominal imaging using 2-dimensional controlled aliasing in parallel imaging results in higher acceleration: a comparison with generalized autocalibrating partially parallel acquisitions parallel imaging.
PURPOSE The purpose of this study was to evaluate the feasibility and technical quality of an abdominal 3-dimensional interpolated breath-hold (volumetric interpolated breath-hold examination [VIBE]) magnetic resonance examination using the new parallel acquisition technique, controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA). MATERIALS AND METHODS In this ins...