Goldmann tonometry tear film error and partial correction with a shaped applanation surface
نویسندگان
چکیده
Purpose The aim of the study was to quantify the isolated tear film adhesion error in a Goldmann applanation tonometer (GAT) prism and in a correcting applanation tonometry surface (CATS) prism. Methods The separation force of a tonometer prism adhered by a tear film to a simulated cornea was measured to quantify an isolated tear film adhesion force. Acrylic hemispheres (7.8 mm radius) used as corneas were lathed over the apical 3.06 mm diameter to simulate full applanation contact with the prism surface for both GAT and CATS prisms. Tear film separation measurements were completed with both an artificial tear and fluorescein solutions as a fluid bridge. The applanation mire thicknesses were measured and correlated with the tear film separation measurements. Human cadaver eyes were used to validate simulated cornea tear film separation measurement differences between the GAT and CATS prisms. Results The CATS prism tear film adhesion error (2.74±0.21 mmHg) was significantly less than the GAT prism (4.57±0.18 mmHg, p<0.001). Tear film adhesion error was independent of applanation mire thickness (R2=0.09, p=0.04). Fluorescein produces more tear film error than artificial tears (+0.51±0.04 mmHg; p<0.001). Cadaver eye validation indicated the CATS prism's tear film adhesion error (1.40±0.51 mmHg) was significantly less than that of the GAT prism (3.30±0.38 mmHg; p=0.002). Conclusion Measured GAT tear film adhesion error is more than previously predicted. A CATS prism significantly reduced tear film adhesion error bŷ41%. Fluorescein solution increases the tear film adhesion compared to artificial tears, while mire thickness has a negligible effect.
منابع مشابه
Goldmann Tonometer Prism with an Optimized Error Correcting Applanation Surface
PURPOSE We evaluate solutions for an applanating surface modification to the Goldmann tonometer prism, which substantially negates the errors due to patient variability in biomechanics. METHODS A modified Goldmann or correcting applanation tonometry surface (CATS) prism is presented which was optimized to minimize the intraocular pressure (IOP) error due to corneal thickness, stiffness, curva...
متن کاملGoldmann tonometer error correcting prism: clinical evaluation
PURPOSE Clinically evaluate a modified applanating surface Goldmann tonometer prism designed to substantially negate errors due to patient variability in biomechanics. METHODS A modified Goldmann prism with a correcting applanation tonometry surface (CATS) was mathematically optimized to minimize the intraocular pressure (IOP) measurement error due to patient variability in corneal thickness,...
متن کاملComparison of different intraocular pressure measurement techniques in normal eyes, post surface and post lamellar refractive surgery
BACKGROUND The purpose of this study was to determine the accuracy of intraocular pressure (IOP) measurement after laser in situ keratomileusis (LASIK) or epithelial laser in situ keratomileusis (Epi)-LASIK using Goldmann applanation tonometry, air puff tonometry, ocular response analyzer corneal compensated IOP (ORA IOPcc) and Pentacam corrected IOP. METHODS A prospective comparative clinica...
متن کاملApplanation pressure function in Goldmann tonometry and its correction.
So far applanation tonometry has not worked out any theoretical basis for correcting the result of intraocular pressure measurement carried out on a cornea with noncalibration dimensions by means of the Goldmann tonometer. All the tables of instrument reading corrections for cornea thickness or cornea curvature radius are based exclusively on measurements. This paper represents an attempt at cr...
متن کاملPressure phosphene tonometry versus Goldmann applanation tonometry for measuring intraocular pressure before and after LASIK.
PURPOSE To compare pressure phosphene tonometry with Goldmann applanation tonometry for measuring intraocular pressure (IOP) before and after LASIK. METHODS Forty-three (18 men and 25 women) consecutive healthy patients underwent complete pre- and postoperative LASIK ophthalmologic assessments including manifest and cycloplegic refraction, keratometry, and central corneal thickness. Three rep...
متن کامل