Early corrosion-related failure of the rejuvenate modular total hip replacement.

نویسندگان

  • Morteza Meftah
  • Amgad M Haleem
  • Matthew B Burn
  • Kevin M Smith
  • Stephen J Incavo
چکیده

BACKGROUND The Rejuvenate modular-neck stem implant (Stryker Orthopaedics, Mahwah, New Jersey) was recently recalled due to corrosion at the femoral neck-stem junction. The purpose of this study was to investigate the rate of corrosion-related failure and survivorship of this implant and analyze the correlation of implant and patient factors with serum metal ion levels and revisions. METHODS From June 2009 to July 2012, 123 Rejuvenate total hip arthroplasty stems (ninety-seven modular and twenty-six non-modular) were implanted in 104 patients by a single surgeon. Serum cobalt (Co) and chromium (Cr) levels (micrograms per liter [μg/L]) were measured postoperatively for all patients. Patients with persistent hip pain or elevated metal ion levels underwent magnetic resonance imaging for assessment of osteolysis or adverse local tissue reactions. Correlation of implant factors (stem size, head size, head length, and femoral head-neck offset) and patient factors (age, sex, and body mass index) with serum metal ion levels and revisions were analyzed with use of logistic regression models. RESULTS The mean duration of follow-up (and standard deviation) was 2.7 ± 0.6 years. The mean Co and Cr levels were 5.4 ± 5.7 μg/L (range, 0.2 to 31 μg/L) and 2.1 ± 1.5 μg/L (range, 0.1 to 4.3 μg/L), respectively. The differences in Co and Cr levels between the two groups (modular and non-modular) were significant: 48% of the total hip arthroplasties in the modular group resulted in elevated metal ion levels (Co >4.0 μg/L and Cr >2.0 μg/L; p < 0.05). The metal ion levels in the non-modular group were normal. In the modular group, higher metal ion levels were significantly correlated with younger age and a higher femoral head-neck offset (p = 0.04). Pain and high Co serum levels were significant predictors of revision surgery (p = 0.006). The rate of revision at the time of this study was 28% in the modular group, with the majority of the revisions performed in the second year after surgery; the Kaplan-Meier survivorship was 40% at four years. CONCLUSIONS The short-term high rate of corrosion-related revision with Rejuvenate modular-neck stems is striking.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Failure of Emperion modular femoral stem with implant analysis

Modularity in total hip arthroplasty provides multiple benefits to the surgeon in restoring the appropriate alignment and position to a previously damaged hip joint. The vast majority of modern implants incorporate modularity into their design with some implants having multiple modular interfaces. There is the potential for failure at modular junctions because of fretting and crevice corrosion ...

متن کامل

Modular titanium alloy neck failure in total hip replacement: analysis of a relapse case.

Modular neck hip prosthesis born in the 1990 with the aim of allowing the surgeon to modify CCD angle, offset and femoral anteversion intra-operatively restoring patient's original biomechanics. In order to achieve the best biomechanics of the reconstructed hip, preoperative planning is essential. In the last few years modularity has been questioned and an argument made for the return to mono b...

متن کامل

Custom titanium sleeve for surgical treatment of mechanically assisted crevice corrosion in the well-fixed, noncontemporary stem in total hip arthroplasty

Adverse local tissue reaction associated with total hip replacement may occur when mechanically assisted crevice corrosion occurs at metal-metal modular junctions in which at least one of the components is fabricated from cobalt-chromium alloy. Complete removal of components may be associated with significant morbidity; when components are well fixed and in acceptable position, it may be approp...

متن کامل

Modular titanium alloy neck adapter failures in hip replacement - failure mode analysis and influence of implant material

BACKGROUND Modular neck adapters for hip arthroplasty stems allow the surgeon to modify CCD angle, offset and femoral anteversion intraoperatively. Fretting or crevice corrosion may lead to failure of such a modular device due to high loads or surface contamination inside the modular coupling. Unfortunately we have experienced such a failure of implants and now report our clinical experience wi...

متن کامل

Corrosion on the acetabular liner taper from retrieved modular metal-on-metal total hip replacements.

Eight retrieved metal-on-metal total hip replacements displayed corrosion damage along the cobalt-chromium alloy liner taper junction with the Ti alloy acetabular shell. Scanning electron microscopy indicated the primary mechanism of corrosion to be grain boundary and associated crevice corrosion, which was likely accelerated through mechanical micromotion and galvanic corrosion resulting from ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of bone and joint surgery. American volume

دوره 96 6  شماره 

صفحات  -

تاریخ انتشار 2014