OBJECTIVES The purpose of this in vitro study was to evaluate the accuracy of working length (WL) determination of four electronic apex locators (EALs), namely, Root ZX (RZX), Elements diagnostic unit and apex locator (ELE), SybronEndo Mini Apex locator (MINI) and Propex pixi (PIXI) using Stainless steel (SS) and nickel-titanium (NiTi) hand files. The null hypothesis was that there was no difference between canal length determination by SS and NiTi files of 4 EALs. MATERIALS AND METHODS Sixty extracted, single rooted human teeth were decoronated and the canal orifice flared. The actual length (AL) was assessed visually, and the teeth were embedded in an alginate model. The electronic length (EL) measurements were recorded with all four EALs using SS and NiTi files at '0.5' reading on display. The differences between the AL and EL were compared. RESULTS The results obtained with each EAL with SS and NiTi files were compared with AL. A paired sample t test showed that there was a statistical significant difference between EAL readings with SS and NiTi files for RZX and MINI (p < 0.05). The accuracy of RZX, ELE, MINI and PIXI within ± 0.5 mm of AL with SS/NiTi files were 93.3%/70%, 90%/91.7%, 95%/68.3%, and 83.3%/83.3%, respectively. CONCLUSIONS The results of this study indicate that Root ZX was statistically more accurate with NiTi files compared to SS files, while MINI was statistically more accurate with SS files compared to NiTi files. ELE and PIXI were not affected by the alloy type of the file used to determine WL.
منابع مشابه
The accuracy of the Raypex5 electronic apex locator using stainless-steel hand K-file versus nickel-titanium rotary Mtwo file.
OBJECTIVES Today many clinicians use both stainless-steel hand K-files and nickel- titanium rotary files during endodontic treatment. It is of great importance for the clinician to have confidence in the accuracy of an apex locator even if these files were used interchangeably. The purpose of this in vitro study was to evaluate the accuracy of the Raypex5 electronic apex locator using stainless...
متن کاملThe accuracy of the Root ZX electronic apex locator using stainless-steel and nickel-titanium files.
Numerous apex locator studies have been performed. Generally, they use only stainless-steel hand files for testing purposes. Today many clinicians use both stainless-steel and nickel-titanium files during the treatment of a case. Given the widespread use of nickel-titanium files, a comparison of the accuracy in determining length with an apex locator using stainless-steel and nickel-titanium fi...
متن کاملAn in vivo comparison of the Root ZX II, the Apex NRG XFR, and Mini Apex Locator by using rotary nickel-titanium files.
INTRODUCTION The purpose of this study was to compare the accuracy of working length (WL) measurements by using the Root ZX II, Apex NRG XFR, and Mini Apex Locator with rotary nickel-titanium (NiTi) instruments. METHODS Twenty-eight teeth had their WLs determined with each electronic apex locator (EAL) by using 0.04 taper ProFiles sizes 40-20 in a crown-down technique until WL was reached. Fo...
متن کاملIn vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination
In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination Dr. Hasheminia, S.M.* - Dr. Jafari, N.** *Associate Professor of Endodontics Dept., Faculty of Dentistry, Isfahan University of Medical Sciences. ** Dentist. Abstract Background & Aim: Radiography is still the common method of measuring working length in root canal therapy. While, gag refl...
متن کاملAn in-vitro Evaluation of canal transportation and centering ability of two rotary Nickel Titanium systems (Twisted Files and Hyflex files) with conventional stainless Steel hand K-flexofiles by using Spiral Computed Tomography.
BACKGROUND The purpose of this in-vitro study was to compare canal transportation and centering ability of Twisted and Hyflex Rotary Files with stainless steel hand k-flexofiles by using Spiral Computed Tomography. MATERIALS & METHODS A total of 90 freshly extracted human mandibular single rooted Premolar teeth were selected. The crowns were flattened with steel disks and a final dimension of...