Implementation of a quality assurance program for computerized treatment planning systems.
نویسندگان
چکیده
In the present investigation, the necessary tests for implementing a quality assurance program for a commercial treatment planning system (TPS), recently installed at Sao Paulo University School of Medicine Clinicas Hospital-Brazil, was established and performed in accordance with the new IAEA publication TRS 430, and with AAPM Task Group 53. The tests recommended by those documents are classified mainly into acceptance, commissioning (dosimetric and nondosimetric), periodic quality assurance, and patient specific quality assurance tests. The recommendations of both IAEA and AAPM documents are being implemented at the hospital for photon beams produced by two linear accelerators. A Farmer ionization chamber was used in a 30 x 30 x 30 cm3 phantom with a dose rate of 320 monitor unit (MU)/min and 50 MU in the case of the dosimetric tests. The acceptance tests verified hardware, network systems integration, data transfer, and software parameters. The results obtained are in good agreement with the specifications of the manufacturer. For the commissioning dosimetric tests, the absolute dose was measured for simple geometries, such as square and rectangular fields, up to more complex geometries such as off-axis hard wedges and for behavior in the build up region. Results were analysed by the use of confidence limit as proposed by Venselaar et al. [Radio Ther. Oncol. 60, 191-201 (2001)]. Criteria of acceptability had been applied also for the comparison between the values of MU calculated manually and MU generated by TPS. The results of the dosimetric tests show that work can be reduced by choosing to perform only those that are more crucial, such as oblique incidence, shaped fields, hard wedges, and buildup region behavior. Staff experience with the implementation of the quality assurance program for a commercial TPS is extremely useful as part of a training program.
منابع مشابه
Developing a Verification and Training Phantom for Gynecological Brachytherapy System
Introduction Dosimetric accuracy is a major issue in the quality assurance (QA) program for treatment planning systems (TPS). An important contribution to this process has been a proper dosimetry method to guarantee the accuracy of delivered dose to the tumor. In brachytherapy (BT) of gynecological (Gyn) cancer it is usual to insert a combination of tandem and ovoid applicators with a complicat...
متن کاملA novel phantom design for brachytherapy quality assurance
Background: One major challenge in brachytherapy is to verify the accuracy of dose distributions calculated by the treatment planning system. In this project, a new phantom design has been introduced for quality assurance of dose distributions in gynocological (GYN) brachytherapy implants using EBT GafChromic film. Materials and Methods: This phantom has been designed and fabricated from 90 sla...
متن کاملAttitudes and Knowledge of Hormozgan University of Medical Sciences Nurses Regarding the Implementation of Computerized Physician Order Entry
Introduction: Despite potential benefits of CPOE (Computerized Physician Order Entry) systems, recent studies have cast some doubts on their role in reducing errors. CPOE systems with poorly designed interfaces have proven to cause dissatisfaction and introduce new kinds of errors in the ordering process. The main objective of this study is to identify problems related to a CPOE medication syst...
متن کاملAmerican Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: quality assurance for clinical radiotherapy treatment planning.
In recent years, the sophistication and complexity of clinical treatment planning and treatment planning systems has increased significantly, particularly including three-dimensional (3D) treatment planning systems, and the use of conformal treatment planning and delivery techniques. This has led to the need for a comprehensive set of quality assurance (QA) guidelines that can be applied to cli...
متن کاملA quality assurance program for an amorphous silicon electronic portal imaging device using in-house developed phantoms: a method development for dosimetry purposes
Background: Electronic portal imaging devices (EPIDs) play an important role in radiation therapy portal imaging, geometric and dosimetric verifications. A successful utilization of EPIDs for imaging and dosimetric purposes requires a reliable quality control process routine to be carried out regularly. In this study, two in-house phantoms were developed and analyzed for implementation in a qua...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical physics
دوره 34 7 شماره
صفحات -
تاریخ انتشار 2007