Leakage of fluid around endotracheal tube cuffs: a cadaver study
نویسندگان
چکیده
BACKGROUND The aim of the study was to evaluate the leakage of liquid past the cuffs of tracheal tubes in fresh frozen human heads. METHODS Six truncated fresh frozen heads were used and intubated with 8.0 mm endotracheal tubes. The intracuff pressures tested were 30 and 100 cmH2O. Subsequently, 20 ml of each of two oral antiseptic rinses (0.2% chlorhexidine and octenidine [octenidol®, Schülke & Mayr GmbH, Norderstedt, Germany]) was applied for thirty seconds in the mouth. During the trial, leakage of the cuffs was examined. RESULTS The sealing between the tracheal cuff and tracheal wall was leakage-proof for all tested intracuff pressures and all tested antiseptic rinses. However, approximately 5.6 ml and 1.8 ml leaked into the esophagus and remained as a cuff-puddle, respectively. CONCLUSIONS The sealing between an endotracheal tube cuff with an intracuff pressure of 30 cmH2O and the tracheal wall is leakage-proof during oral care with antiseptic rinsing. An increase of intracuff pressure to 100 cmH2O does not appear to be required.
منابع مشابه
A quantitative evaluation of fluid leakage around a polyvinyl chloride tapered endotracheal tube cuff using an in-vitro model
INTRODUCTION This study was designed to quantitatively evaluate the sealing effect of a polyvinyl chloride tapered endotracheal tube cuff and to compare the tapered cuff with cylindrical endotracheal tube cuffs using an in vitro viscous fluid model. METHODS Five types of 8.0 mm inner diameter endotracheal tubes (TaperGuard, PortexSacett, PortexSoftseal, Sheridan HVT, Sheridan CF) were fixed i...
متن کاملAssessment of aspiration risk from dynamic modulation of endotracheal tube cuff pressure.
OBJECTIVES/HYPOTHESIS To assess the risk of aspiration using a novel valve circuit that dynamically modulates endotracheal tube cuff pressure during the ventilatory cycle using bench and live animal models. STUDY DESIGN Animal model. METHODS The bench model consisted of a cuffed endotracheal tube inserted into an artificial trachea. Leakage of liquid around the cuff was measured after 4 hou...
متن کاملAn in vitro microbiological study comparing eight endotracheal tubes and their ability to prevent microaspiration
Introduction The major cause of ventilator-associated pneumonia (VAP) is the aspiration of bacteria-laden subglottic secretions past the cuff of the endotracheal tube (ETT) [1]. When the ETT cuff is inflated to the correct wall pressure, excess cuff material folds and causes involutions thereby forming channels, which allow leakage of subglottic secretions to the lungs [2]. Now, new ETT cuffs h...
متن کاملComparison of a novel lycra endotracheal tube cuff to standard polyvinyl chloride cuff and polyurethane cuff for fluid leak prevention.
BACKGROUND A high-volume low-pressure endotracheal tube (ETT) cuff forms folds along its contact with the trachea, allowing mucus leakage into the lungs. We developed a thin-walled ETT cuff made of Lycra polyurethane. METHODS In vitro, we tested 6 of each of the new prototype Lycra cuff, the Mallinkrodt Hi-Lo ETT (polyvinyl chloride cuff), and the Kimberly-Clark Microcuff ETT (polyurethane cu...
متن کاملThe value of polyurethane-cuffed endotracheal tubes to reduce microaspiration and intubation-related pneumonia: a systematic review of laboratory and clinical studies
BACKGROUND When conventional high-volume, low-pressure cuffs of endotracheal tubes (ETTs) are inflated, channel formation due to folds in the cuff wall can occur. These channels facilitate microaspiration of subglottic secretions, which is the main pathogenic mechanism leading to intubation-related pneumonia. Ultrathin polyurethane (PU)-cuffed ETTs are developed to minimize channel formation in...