Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

نویسندگان

  • Mehenna Arib
  • Toufik Medjadj
  • Youcef Boudouma
چکیده

In the International Atomic Energy Agency's (IAEA) code of practice (TRS 398) and the American Association of Physicists in Medicine's dosimetry protocol (TG-51), full-scatter water phantoms are recommended for the determination of the absorbed dose for both photon and electron beams and, consequently, for the calibration of the user's ionization chambers. This procedure is applied in the Secondary Standard Dosimetry Laboratory, where the calibration is performed on a 60Co gamma beam, in comparison with reference chambers whose absorbed dose-to-water calibration coefficients, ND,w, are known. In this work, we present the results of the calibration of 10 Farmer-like ionization chambers calibrated in three water phantoms (sizes 20 x 20 x 15 cm3, 30 x 30 x 30 cm3, and 35 x 35 x 37 cm3) and two plastic phantoms (size 20 x 20 x 20 cm3) polymethyl methacrlyate (PMMA) and polystyrene). Calibrations are performed by the substitution method using an ionization chamber whose ND,w has been supplied by the IAEA's reference laboratory. It is shown that the results, expressed as the percentage ratio of the calibration coefficient in a given phantom to that of the standard IAEA phantom, is less than 0.35% for all investigated chambers, and that the standard deviation of the mean of the ND,w calibration coefficients determined in all five phantoms is less than 0.06%, except for one nylon-walled ionization chamber, where the observed 0.34% value could be explained by the hygroscopic properties of nylon. Furthermore, a chamber-to-chamber dependence of the calibration coefficient has been shown to vary by up to 2.8%. These results emphasize that the phantom dimensions and its material are not sensitive criteria for the calibration of cylindrical ionization chambers in terms of absorbed dose to water. The results also show that generic calibration coefficients could not be considered for a given type of chamber.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Background: Absolute dosimetry of external beam radiotherapy is carried out by the use of ionization chambers. These chambers must be calibrated at a standard dosimetry laboratory before any use in clinical dosimetry. The secondary standard dosimetry laboratory of Iran (SSDL) has the duty of calibrating the ionization chambers used in radiotherapy centers in Iran. Materials and Methods...

متن کامل

The effect of electronic disequilibrium on the received dose by lung in small fields with photon beams: Measurements and Monte Carlo study

Background: Prediction of the absorbed dose in irradiated volume plays an important role in the outcome of radiotherapy. Application of small fields for radiotherapy of thorax makes the dose calculation process inaccurate due to the existence of electronic disequilibrium and intrinsic deficiencies in dose calculation algorithms. To study the lung absorbed dose in radiotherapy with smal...

متن کامل

Absorbed dose to water reference dosimetry using solid phantoms in the context of absorbed-dose protocols.

For reasons of phantom material reproducibility, the absorbed dose protocols of the American Association of Physicists in Medicine (AAPM) (TG-51) and the International Atomic Energy Agency (IAEA) (TRS-398) have made the use of liquid water as a phantom material for reference dosimetry mandatory. In this work we provide a formal framework for the measurement of absorbed dose to water using ioniz...

متن کامل

Some steps towards establishing a tertiary standard dosimetry laboratory at a radiotherapy department

Background: In order to deliver the precise dose to the target in radiotherapy, absorbed dose to water at the reference point should be assessed. When the calibration procedure is performed for a reference dosimeter in the 60Co beam of a Secondary Standard Dosimetry Laboratory (SSDL), the total uncertainty in absorbed dose to water (Dw) is estimated to be approximately 1.5%. This study attempts...

متن کامل

Calculations of absorbed dose and energy dependent of small-scale dosimeters for photons beam therapy

In this study, the energy dependency for dosimeters of air ionization chambers, lithium fluoride, silicon and plastic scintillator has been studied using the MCNPX Monte Carlo simulation code and simulated for gamma energy in the range of radiotherapy energy. The simulation results show that the response of each of the dosimeters for gamma photon beams in the energy range of 0.2 to 20 MeV varie...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 7  شماره 

صفحات  -

تاریخ انتشار 2006