Volume guarantee ventilation, interrupted expiration, and expiratory braking.
نویسندگان
چکیده
BACKGROUND In volume guarantee ventilation with the Dräger Babylog 8000 ventilator, inspiratory and expiratory flows are monitored and the expiratory tidal volume calculated following each inflation. The pressure for the next inflation is modified to ensure the expired tidal volume is close to the set value. AIM To investigate interrupted expiration observed during volume guarantee ventilation of spontaneously breathing, ventilated infants. METHODS Spontaneously breathing infants, ventilated with volume guarantee, had recordings of gas flow, ventilator pressures, tidal volume waveforms, oximetry, heart rate, and transcutaneous oxygen and carbon dioxide during 10 minute recordings. RESULTS A total of 6540 inflations were analysed from 10 infants; 62% were triggered. Two different patterns were found: (1) Normal volume guarantee pattern with 97% of triggered and 91% untriggered inflations. It had a normal expiratory curve and a mean expired tidal volume within 3% of the set volume, but a large variation due to the babies' breathing. (2) A pattern of interrupted expiratory flow after approximately 3% of inflations due to a small inspiration (approximately 1.3 ml/kg) during expiration. This led the ventilator to calculate an inappropriate total expired tidal volume for that inflation and an increase in the pressure for the next inflation. CONCLUSIONS After about 3% of inflations, with volume guarantee ventilation, interruption of the expiration causes an increased pressure for the next inflation of approximately 4.9 cm H2O, compared with normal volume guarantee inflation. The interrupted expiration is most likely to be due to diaphragmatic braking.
منابع مشابه
Breathing muscle activity during expiration in patients with chronic airflow obstruction.
During quiet ventilation in 10 patients with severe chronic airflow obstruction (AFO) there were large tidal swings of pleural pressure, 15.8 +/- 5.1 cmH2O, with high negative pressures achieved, 15.2 +/- 5.2 cmH2O. The pattern of pleural pressure mirrored that of tidal expiratory flow with a rapid rise to maximum of pleural pressure and flow at the beginning of expiration, a slow decline of bo...
متن کاملThe diaphragm has an expiratory braking effect in spontaneously breathing lung injured animals as shown by electrical diaphragmatic activity
Introduction Albeit defined as an inspiratory muscle, the diaphragm interacts with the respiratory system also during expiration. The expiratory load on the diaphragm derives from lung volume and hence, possibly from lung collapse. We hypothesized that the diaphragm has a braking effect during the whole expiration, preserving lung patency during SB in collapse prone lungs. We investigated this ...
متن کاملParameters affecting the tidal volume during expiratory abdominal compression in patients with prolonged tracheostomy mechanical ventilation
[Purpose] The aim of this study was to clarify physical parameters affecting the tidal volume during expiratory abdominal compression in patients with prolonged tracheostomy mechanical ventilation. [Methods] Eighteen patients with prolonged mechanical ventilation were included in this study. Expiratory abdominal compression was performed on patients lying in a supine position. The abdomen above...
متن کاملTime constant/volume relationship of passive expiration in mechanically ventilated ARDS patients.
Since the adult respiratory distress syndrome (ARDS) lung is known to be inhomogeneous, one could expect an uneven distribution of expiratory time constant during uninterrupted mechanical ventilation. We investigated the time constant/volume relationship of passive expiration, and their modification by external resistive elements. In 12 paralysed intubated ARDS patients, we determined the expir...
متن کاملComparison of changes in tidal volume associated with expiratory rib cage compression and expiratory abdominal compression in patients on prolonged mechanical ventilation
[Purpose] This study was designed to compare and clarify the relationship between expiratory rib cage compression and expiratory abdominal compression in patients on prolonged mechanical ventilation, with a focus on tidal volume. [Subjects and Methods] The subjects were 18 patients on prolonged mechanical ventilation, who had undergone tracheostomy. Each patient received expiratory rib cage com...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Archives of disease in childhood
دوره 90 8 شماره
صفحات -
تاریخ انتشار 2005