Bromide Determination in Pharmaceutical Preparations Using Indirect Gas-Diffusion Flow Injection Method with Amperometric Detector
نویسنده
چکیده
The indirect gas-diffusion flow injection method (FI) was modified, optimized and successfully applied to bromide determination in various samples such as active substances, drugs, organic substances and water samples. The method is based on the oxidation of bromide to bromine in acidic medium, which diffuses through PTFe membrane into the acceptor flow that carried bromine to the amperometric cell. The flow amperometric detection was carried out at the potential of +0.65 V (vs Ag/AgCl) when injected sample volume was 100μl. results obtained by FIA were confirmed by cyclic voltammetry (CV). The linear calibration graph was obtained over the concentration range 0.1–10x10-3 mol/l and 2.5–100x10-6 mol/l bromide solutions. relative standard deviation was 1.23% for 5x10-6 mol/l bromide solution and 0.11% for 0.25x10-3 mol/l bromide solution (n=6, for both cases). The limit of detection was 5x10-8 mol/l. The described method requires small quantities of samples and prior extraction of the analytes is not necessary. The method enables 60 analyses per hour.
منابع مشابه
Determination of Vitamin C (Ascorbic Acid) Using High Performance Liquid Chromatography Coupled with Electrochemical Detection
Vitamin C (ascorbic acid, ascorbate, AA) is a water soluble organic compound that participates in many biological processes. The main aim of this paper was to utilize two electrochemical detectors (amperometric - Coulouchem III and coulometric - CoulArray) coupled with flow injection analysis for the detection of ascorbic acid. Primarily, we optimized the experimental conditions. The optimized ...
متن کاملRapid High Performance Liquid Chromatographic Method for Determination of Clarithromycin in Human Plasma Using Amperometric Detection: Application in Pharmacokinetic and Bioequivalence Studies
A rapid, sensitive and reproducible HPLC method using amperometric detector was developed and validated for the analysis of clarithromycin in human plasma. The separation was achieved on a monolithic silica column (MZ- C8 125×4.0 mm) using acetonitrile- methanol-potassium dihydrogen phosphate buffer (40:6:54,v/v), with pH of 7.5, as the mobile phase at a flow rate of 1.5 mL/min. The assay enabl...
متن کاملFlow-Injection Amperometric Determination of Ascorbic Acid Using a Graphite-Epoxy Composite Electrode Modified with Cobalt Phthalocyanine
A flow injection method is reported for the determination of ascorbic acid based on amperometric detection, using a cobalt(II) phthalocyanine (CoPc) modified graphite-epoxy electrode. The amperometric respons was evaluated with regard to pH, ionic strength of the electrolyte, flow rate of the carrier solution, injected sample volume and conditioning time of the electrode. The limit of detec...
متن کاملRapid High Performance Liquid Chromatographic Method for Determination of Clarithromycin in Human Plasma Using Amperometric Detection: Application in Pharmacokinetic and Bioequivalence Studies
A rapid, sensitive and reproducible HPLC method using amperometric detector was developed and validated for the analysis of clarithromycin in human plasma. The separation was achieved on a monolithic silica column (MZ- C8 125×4.0 mm) using acetonitrile- methanol-potassium dihydrogen phosphate buffer (40:6:54,v/v), with pH of 7.5, as the mobile phase at a flow rate of 1.5 mL/min. The assay enabl...
متن کاملDifferential electrolytic potentiometry, a detector in flow injection analysis for oxidation-reduction reactions.
For the first time, differential electropotentiometry (DEP) is coupled with the flow injection analysis (FIA) technique for detection of oxidation-reduction reactions, and is utilized for quantitative determination of vitamin C in pharmaceutical preparations using 1.0x10(-3)-M cerium(IV) in 0.50-M sulfuric acid as carrier. Two similar platinum electrodes were employed and polarized by a constan...