Augmented glenoid component designs for type B2 erosions: a computational comparison by volume of bone removal and quality of remaining bone.

نویسندگان

  • Nikolas K Knowles
  • Louis M Ferreira
  • George S Athwal
چکیده

BACKGROUND The purpose of this computational modeling study was to compare the volume of glenoid bone removal required to implant 3 augmented component designs for management of B2 erosions. In addition, we assessed bone quality of the supporting bone directly beneath the implants by measuring bone density and porosity. METHODS Three augmented component designs—full-wedge, posterior-wedge, and posterior-step—were studied by virtual implantation in a cohort of 16 patients with B2 glenoids. B2 retroversion was corrected to 0° and 10°. The outcome variables were the volume of glenoid bone removal required for implantation and the density and porosity of the bone immediately beneath the implant. RESULTS Implant design had a significant effect on the volume of bone removal (P < .001). When correcting to 0°, the posterior-wedge implant removed less bone than the posterior-step (P < .001) and the full-wedge (P = .004). At 10° retroversion, the posterior-wedge removed less bone (P = .029) than the posterior-step but was no different than the full-wedge (P = .143). The residual glenoid bone density with the posterior-wedge was significantly greater than with the posterior-step (P = .048), with no other significant differences (P > .05). Residual glenoid bone porosity was not significantly different between implants (P > .262). CONCLUSIONS Augmented components can provide a bone-preserving option for B2 glenoid management. Substantial variations in the volume of bone removal and the quality of the remaining glenoid bone were found between 3 different designs of augmented implants. Simulations with the posterior-wedge implant resulted in substantially less glenoid bone removal, with the remaining supporting bone being of better quality.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of Asymmetric Reaming versus a Posteriorly Augmented Component for Posterior Glenoid Wear and Retroversion: A Radiographic Study

Background: Managing posterior glenoid wear and retroversion remains a challenge in shoulder arthroplasty.Correcting glenoid version through asymmetric reaming (AR) with placement of a standard glenoid component and theuse of posteriorly augmented glenoid (PAG) components are two methods used to address this problem. Our objectiveis to report the radiographic outcomes of patie...

متن کامل

Determination of prevalence of glenoid bony lesions after recurrent anterior shoulder dislocation using the 3-D CT scan

  Background : Glenoid bone damages consisting of anterior rim erosion and bony avulsion are very important in decision making for treatment of recurrent dislocation in shoulder joint. This study was aimed to determine the prevalence of these damages in patients with anterior recurrent shoulder dislocation.   Methods : The study was a cross-sectional study evaluating patients with unstable shou...

متن کامل

Results of Open Bankart Surgery for Recurrent Anterior Shoulder Dislocation with Glenoid Bone Defect and Concomitant Hill-Sachs Lesion

  Background: Open Bankart surgery is the main treatment procedure in patients with recurrent anterior shoulder dislocation, especially in cases with glenoid bone defect. The goal of this study was to determine the outcomes after open Bankart surgery in cases of recurrent anterior shoulder dislocation with glenoid bone defects and concomitant Hill-Sachs lesions. Methods: Between 2006 and 2010,...

متن کامل

Glenoid bone loss in primary total shoulder arthroplasty: evaluation and management.

Glenohumeral osteoarthritis is the most common reason for shoulder replacement. Total shoulder arthroplasty provides reliable pain relief and restoration of function, with implant survivorship reported at 85% at 15 years. Glenoid component wear and aseptic loosening are among the most common reasons for revision. Glenoid wear characteristics have been correlated with, among other things, the de...

متن کامل

Edge displacement and deformation of glenoid components in response to eccentric loading. The effect of preparation of the glenoid bone.

The effect of different methods of preparation of glenoid bone on displacement and deformation of a glenoid component under eccentric loading was investigated in a series of scapulae of cadavera. Handburring of the osseous surface was associated with less displacement and deformation than simple removal of cartilage with a curet. Reaming resulted in the least displacement and deformation. Subst...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of shoulder and elbow surgery

دوره 24 8  شماره 

صفحات  -

تاریخ انتشار 2015