Complete Source Trajectories for C-Arm Systems and a Method for Coping with Truncated Cone-Beam Projections
نویسنده
چکیده
1 Introduction A C-arm system, such as the one shown in Figure 1, may be used to acquire X-ray cone-beam projections of a patient's three-dimensional (3D) region of interest (ROI) while the X-ray source moves along some trajectory around the ROI. Then, a 3D image of the X-ray attenuation coefficient within the ROI may be reconstructed from the acquired cone-beam projections. In fact, C-arm systems are being used in this way, see e.g. [1, 2]. Typically, the source moves along a circular arc spanning an angle of 180 • –200 • , and the image is reconstructed by a variant of the algorithm of Feldkamp, Davis, and Kress [3]. As the cone-beam does not cover the whole patient, the projections are truncated. This is usually handled by extending the projections in a simple manner parallel to the plane containing the source trajectory. In practice, the true trajectory differs slightly from its ideal, but this deviation can be measured [5, 6] and taken into account during the reconstruction. The reconstructed image is subjected to a surface or volume rendering process designed to extract and visualize only the high contrast structures of the object under examination. The reconstructed image itself is usually cluttered by severe artifacts. Nevertheless, the high contrast structures, such as bones or blood vessels filled with intraarterially injected contrast agents, are well recovered. This is because such structures stand out well against the background and are also reconstructed at the correct geometric locations. Accurate images of medium contrast structures, such as blood vessels filled with intravenously injected contrast agents, or low contrast structures, such as soft tissue organs, are not obtained in this way. If this is to be improved upon, the following conditions will have to be met: First, the data acquired by the C-arm system must provide (after some preprocessing) accurate, though sampled, cone-beam projections of the object function (the X-ray attenuation coefficient). Second, the sampling density along the trajectory and on the detector surface must be sufficiently high. Third, the source trajectory must be complete in the sense that every plane that intersects the ROI contains a source point. Fourth, the cone beam projections must not be truncated. Under these conditions, any standard exact cone-beam reconstruction algorithm will produce an accurate estimate of the object function within the ROI. It has been tacitly assumed that the object function varies only spatially. Imaging moving parts of the …
منابع مشابه
2D filtered backprojection for fan-beam CT with independent rotations of the source and the detector
A cone-beam CT scanner has recently been developed for radiotherapy imaging where the source and the flat detector can rotate independently along concentric circular trajectories. This paper investigates the reconstruction of the central slice of this system. A new filtered-backprojection algorithm has been derived that only modifies the weighting schemes of the projections and the backprojecti...
متن کاملI. INTRODUCTION Advances in model-based iterative reconstruction (IR) methods for x-ray CT and cone-beam CT (CBCT) imaging
C-arm cone-beam CT offers great potential in image-guided interventions, but conventional analytic reconstruction methods are associated with limited image quality, particularly for soft-tissue imaging. While model-based iterative reconstruction (IR) methods improve image quality and/or reduce radiation dose, long reconstruction time limits utility in clinical workflow. Additionally, in contras...
متن کاملMega Voltage Cone Beam Computed Tomography (MV- CBCT) using a Standard Medical Linear Accelerator and EPID: A feasibility study.
Introduction: The success of radiotherapy cancer treatment delivery depends on the accuracy of patient positioning for each treatment session. A number of kilovoltage x-ray volumetric imaging modalities with an additional source and detector have been developed to allow patient set-up verification based on the internal anatomy, but a significant portion of medical linacs are on...
متن کاملRoi Cone-beam Ct on a Circular Orbit for Geometric Magnification Using Reprojection
We present an algorithm which allows us to investigate what is feasible when reconstructing a Region-Of-Interest (ROI) from transaxial truncated cone-beam projections obtained on a circular orbit in combination with information obtained at a different magnification level. We use the well-known fact that a rough estimate of the global properties of the Radon Transform is sufficient to obtain a r...
متن کاملLine plus arc source trajectories and their R-line coverage for long-object cone-beam imaging with a C-arm system.
Cone-beam imaging with C-arm systems has become a valuable tool in interventional radiology. Currently, a simple circular trajectory is used, but future applications should use more sophisticated source trajectories, not only to avoid cone-beam artifacts but also to allow extended volume imaging. One attractive strategy to achieve these two goals is to use a source trajectory that consists of t...
متن کامل