FMECA Application to Intraoperative Electron Beam Radiotherapy Procedure As a Quality Method to Prevent and Reduce Patient’s Risk in Conservative Surgery for Breast Cancer
نویسندگان
چکیده
INTRODUCTION Failure Mode Effects and Criticalities Analysis (FMECA) represents a prospective method for risk assessment in complex medical practices. Our objective was to describe the application of FMECA approach to intraoperative electron beam radiotherapy (IOERT), delivered using a mobile linear accelerator, for the treatment of early breast cancer as an anticipated boost. MATERIALS AND METHODS A multidisciplinary Working Group, including several different professional profiles, was created before the beginning of clinical practice in 2012, with the purpose of writing the Flow Chart and applying the FMECA methodology to IOERT procedure. Several criticalities were identified a priori in the different steps of the procedure and a list of all potential failure modes (FMs) was drafted and ranked using the risk priority number (RPN) scoring system, based on the product of three parameters: severity, occurrence, and detectability (score between 1 and 5). The actions aimed at reducing the risk were then defined by the Working Group and the risk analysis was repeated in 2014 and in 2016, in order to assess the improvement achieved. RESULTS Fifty-one FMs were identified, which represented the issues prospectively investigated according to the FMECA methodology. Considering a set threshold of 30, the evaluated RPNs show that 33 out of 51 FMs are critical; 6 are included in the moderate risk class (RPN: 31-40); 16 in the intermediate risk class (RPN: 41-50), and 11 in the high risk class (RPN: >50). DISCUSSION The most critical steps concerned the surgical procedure and IOERT set-up. The introduction of the corrective actions into the clinical practice achieved the reduction of the RPNs in the re-analysis of the FMECA worksheet after 2 and 4 years, respectively. CONCLUSION FMECA proved to be a useful tool for prospective evaluation of potential failures in IOERT and contributed to optimize patient safety and to improve risk management culture among all the professionals of the Working Group.
منابع مشابه
Dosimetric pitfalls in the application of shielding disk in breast intraoperative electron radiation therapy
Introduction: Intraoperative electron radiotherapy (IOERT) is an IORT technique in which electron beams with different nominal energies are used for irradiation of microscopic residual of tumor bed after surgery. Application of a shielding disk, which is positioned under the target volume, for protecting the normal tissues such as pectoral muscle, lung, and heart is one of the a...
متن کاملCommissioning the First Mobile Dedicated Accelerator for Intraoperative Electron Radiotherapy in Iran
Introduction: Intraoperative radiotherapy is a radiotherapy technique in which a high single fraction of radiation dose is delivered to the patient after surgery and Concurrent with anesthesia time. The most frequent method for IORT implementation is Intraoperative electron radiotherapy (IOERT), in which, some dedicated and high dose per pulse electron accelerators are employe...
متن کاملDosimetric Evaluation of Target Volume in Breast Boost Radiotherapy: Comparison of Electron and Photon Beam
Background and purpose: Breast‑conserving surgery followed by radiation therapy to the whole breast is now recognized as a standard strategy in patients with breast cancer. Recommended technique for radiotherapy is whole breast irradiation followed by boost to the tumor bed. This study aimed to compare the dosimetric parameters of electron and photon beams for boosting irradiation in post‑lump...
متن کاملIntraoperative radiotherapy (IORT) induced key molecular pathways in tumor bed of breast cancer patients: a pilot study
Background: Radiotherapy (RT) is recommended to all patients undergoing Breast Conserving Surgery (BCS). Two strategies can be applied to irradiation, External Beam RT (EBRT) in addition, Intraoperative Radiation Therapy (IORT). The aim of this study was to introduce a protein biomarker panel related to molecular function of IORT. Materials and Methods: Six Breast Cancer (BC) patients as a pilo...
متن کاملشبیهسازی شتاب دهنده پرتو درمانی حین عمل LIAC به همراه اپلیکاتور شکل دهنده باریکه بااستفاده از محاسبات مونتکارلو
Intraoperative electron radiotherapy is one of the radiotherapy methods that delivers a high single fraction of radiation dose to the patient in one session during the surgery. Beam shaper applicator is one of the applicators that is recently employed with this radiotherapy method. This applicator has a considerable application in treatment of large tumors. In this study, the dosimetric charact...
متن کامل