Evaluation of the THRESHOLD trainer for inspiratory muscle endurance training: comparison with the weighted plunger method.

نویسندگان

  • P H Johnson
  • A J Cowley
  • W J Kinnear
چکیده

Inspiratory muscle training (IMT) has been shown to enhance exercise performance. The weighted plunger (WP) system of inspiratory threshold loading is the most commonly used method of IMT, but is expensive and cumbersome. We have evaluated a commercially available portable spring-loaded IMT device, the THRESHOLD trainer. The WP and THRESHOLD trainer devices were evaluated with their opening pressures set, in random order, at 10, 20, 30 and 40 cmH2O. Using an airpump, pressure at the valve inlet was recorded at the point at which the valve opened, and at airflow rates of 20, 40, 60, 80 and 100 L.min-1. Ten THRESHOLD trainers were then compared using the same opening pressures and airflow rates. Finally, 10 patients with stable chronic heart failure (CHF) inspired, in random order, through the WP and THRESHOLD trainer for 4 min each. The pressure-time product (PTP) was calculated for each 4 min period, to compare the work performed on inspiring through each device. The mean measured opening pressures for the WP set at 10, 20, 30 and 40 cmH2O, were 9.0, 19.3, 27.9 and 39.2 cmH2O, respectively, and there was little change over the range of flow tested. Corresponding values for the THRESHOLD trainer were 7.5, 16.9, 26.2 and 39.1 cmH2O, with the pressure being closer to the set pressure as flow increased to that seen in clinical practice. The 10 different trainers tested performed very similarly to one another. Work performed (as measured by PTP) on inspiring through the WP and THRESHOLD trainer was not significantly different. Although less accurate than the weighted plunger, the THRESHOLD trainer is an inexpensive device of consistent quality. In a clinical setting it would be a satisfactory option for inspiratory muscle training in most patients, but less so in patients with very low inspiratory flow rates.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A threshold loading device for testing of inspiratory muscle performance.

Threshold loading is a method of incrementally loading the inspiratory muscles, which requires development of an adjustable threshold pressure (Pth) before air will flow. In the ideal device, flow is independent of pressure once Pth is exceeded. A "weighted plunger" apparatus is commonly used, but devices described previously perform unreliably at high inspiratory load. We have modified this ap...

متن کامل

A Review of Effects of Inspiratory Muscle Training on Clinical and Functional Outcomes of Patients with Mechanical Ventilation

Background and Objectives: This review aims to explore the use of inspiratory muscle training (IMT) in patients with mechanical ventilation (MV). The topics were related to its effect on the duration of MV or weaning, respiratory symptoms or lung function, inspiratory muscle strength (IMS) or endurance, functional ability, and quality of life (QoL).  Methods: Articles published in the last 10 ...

متن کامل

Inspiratory muscle training in quadriplegic patients.

OBJECTIVE To determine whether inspiratory muscle training can increase strength and endurance of these muscles in quadriplegic patients. METHODS Eight quadriplegic patients (7 males and 1 female) with injury to the lower cervical spine (segments C4-C7) were submitted to inspiratory muscle training with a threshold inspiratory muscle trainer adjusted to 30% of MIP. The training sessions were ...

متن کامل

Measurement of inspiratory muscle performance with incremental threshold loading: a comparison of two techniques.

BACKGROUND Incremental threshold loading (ITL) is a test of inspiratory muscle performance which is usually performed by breathing through a weighted inspiratory plunger, the load on the inspiratory muscles being increased by externally adding weights to the intake valve. This is not a true threshold device and may be inaccurate. This method was compared with a true threshold device consisting ...

متن کامل

Comparing the Effect of Resistive Inspiratory Muscle Training and Incentive Spirometry on Respiratory Pattern of COPD patients

Background: Resistive Inspiratory Muscle Training (RIMT) is a well-known technique for rehabilitation of patients with Chronic Obstructive Pulmonary Disease (COPD). Incentive spirometry is another technique with potential viability for this application, but there is limited evidence in support of its efficacy in the rehabilitation of COPD patients. Aim: The objective of this study was to compar...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The European respiratory journal

دوره 9 12  شماره 

صفحات  -

تاریخ انتشار 1996