Quantitative measurement of porphobilinogen in urine by stable-isotope dilution liquid chromatography-tandem mass spectrometry.
نویسندگان
چکیده
BACKGROUND Measurement of porphobilinogen (PBG) is useful in the diagnosis of the acute neurologic porphyrias. Currently used colorimetric assays lack analytical and clinical sensitivity and specificity. METHODS We developed a liquid chromatography-electrospray tandem mass spectrometry (LC-MS/MS) method for the measurement of PBG in 1 mL of urine, using 5-(aminoethyl)-4-(carboxymethyl) 1H-2,4-[(13)C]pyrolle-3-propanoic acid ([2,4-(13)C]PBG; 2.75 microg) as internal standard. After solid-phase extraction, LC-MS/MS analysis was performed in the selected-reaction monitoring (SRM) mode. PBG and [2,4-(13)C]PBG were monitored through their own precursor and product ion settings (m/z 227 to 210 and m/z 229 to 212, respectively). The retention time of PBG and [2,4-(13)C]PBG was 1.0 min in a 2.3-min analysis. RESULTS Daily calibrations (n = 6) between 0.1 and 2.0 mg/L were linear and reproducible. Inter- and intraassay CVs were 3.2-3.5% and 2.6-3.1%, respectively, at mean concentrations of 0.24, 1.18, and 2.15 mg/L. The regression equation for the comparison between an anion-exchange column method (y) and the LC-MS/MS method (x) was: y = 0.84x + 0.74 (S(y/x) = 5.8 mg/24 h; r = 0.85; n = 100). In 47 volunteers, PBG excretion was 0.02-0.42 mg/24 h, lower than reported reference intervals (up to 2.0 mg/24 h) based on colorimetric methods. In 85 samples with PBG < or =0.5 by LC-MS/MS, 8 (9.4%) had values > or =2.0 mg/24 h by the anion-exchange method (mean +/- SD, 4.3 +/- 1.8 mg/24 h). In 11 patients with confirmed diagnoses of acute porphyria and increased PBG by LC-MS/MS, 2 had values within the reported reference intervals by a quantitative anion-exchange method. CONCLUSIONS The quantitative LC-MS/MS method for PBG measurement exhibits greater analytical specificity and improved clinical sensitivity and specificity than currently available methods.
منابع مشابه
Relative Determination Approach to the Metabolites of Protoberberine Alkaloids in Rat Urine by Liquid Chromatography Tandem Mass Spectrometry for the Comparative Studies on Rhizome coptidis and Zuojinwan Preparation
The lack of authentic standards has limited the quantitative analysis of herbal drugs in biological samples. The present work demonstrated a practicable strategy for the assay of herbs and their metabolites independent of authentic standards. A liquid chromatography–electrospray ionization–mass spectrometry (LC–ESI–MS) method for the qualitative and quantitative determination of the metabolites...
متن کاملRelative Determination Approach to the Metabolites of Protoberberine Alkaloids in Rat Urine by Liquid Chromatography Tandem Mass Spectrometry for the Comparative Studies on Rhizome coptidis and Zuojinwan Preparation
The lack of authentic standards has limited the quantitative analysis of herbal drugs in biological samples. The present work demonstrated a practicable strategy for the assay of herbs and their metabolites independent of authentic standards. A liquid chromatography–electrospray ionization–mass spectrometry (LC–ESI–MS) method for the qualitative and quantitative determination of the metabolites...
متن کاملQuantitative detection of trichloroacetic acid in human urine using isotope dilution high-performance liquid chromatography-electrospray ionization tandem mass spectrometry.
The chemical disinfection of drinking water to control microbial contaminants results in the formation of disinfection byproducts (DBPs). The volatile trihalomethanes and the nonvolatile haloacetic acids (HAAs) are the most prevalent DBPs. It is important to monitor human exposure to HAAs because of their potential adversehealth effects, such as cancer. Among the HAAs, urinary trichloroacetic a...
متن کاملDetermination of creatinine in human urine with flow injection tandem mass spectrometry.
BACKGROUND/AIMS Excretion of urinary compounds in spot urine is often estimated relative to creatinine. For the growing number of liquid chromatography-tandem mass spectrometry (LC-MS/MS) assays of urine-excreted molecules, a fast and accurate method for determination of creatinine is needed. METHODS A high-throughput flow injection tandem mass spectrometry method for exact quantitation of cr...
متن کاملDevelopment of a candidate reference method for determination of tyrosine in serum by isotope dilution liquid chromatography-tandem mass spectrometry.
An isotope dilution liquid chromatography/tandem mass spectrometry is proposed as a reference method to determine the level of tyrosine in human serum. The advantages of this method include simple sample preparation without derivatization, the selective detection of compounds of interest in complex matrices, and the use of an isotopically labeled analogue as an internal standard. Tyrosine and i...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 47 9 شماره
صفحات -
تاریخ انتشار 2001