Pulse oximeter sensor application during neonatal resuscitation: a randomized controlled trial.

نویسندگان

  • Deepak Louis
  • Venkataseshan Sundaram
  • Praveen Kumar
چکیده

OBJECTIVE This study was done to compare 2 techniques of pulse oximeter sensor application during neonatal resuscitation for faster signal detection. METHODS Sensor to infant first (STIF) and then to oximeter was compared with sensor to oximeter first (STOF) and then to infant in ≥28 weeks gestations. The primary outcome was time from completion of sensor application to reliable signal, defined as stable display of heart rate and saturation. Time from birth to sensor application, time taken for sensor application, time from birth to reliable signal, and need to reapply sensor were secondary outcomes. An intention-to-treat analysis was done, and subgroup analysis was done for gestation and need for resuscitation. RESULTS One hundred fifty neonates were randomized with 75 to each technique. The median (IQR) time from sensor application to detection of reliable signal was longer in STIF group compared with STOF group (16 [15-17] vs. 10 [6-18] seconds; P <0.001). Time taken for application of sensor was longer with STIF technique than with STOF technique (12 [10-16] vs. 11 [9-15] seconds; P = 0.04). Time from birth to reliable signal did not differ between the 2 methods (STIF: 61 [52-76] seconds; STOF: 58 [47-73] seconds [P = .09]). Time taken for signal acquisition was longer with STIF than with STOF in both subgroups. CONCLUSIONS In the delivery room setting, the STOF method recognized saturation and heart rate faster than the STIF method. The time from birth to reliable signal was similar with the 2 methods.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Pulse Oximeter Sensor Application During Neonatal Resuscitation: A Randomized Controlled Trial

OBJECTIVE: This study was done to compare 2 techniques of pulse oximeter sensor application during neonatal resuscitation for faster signal detection. METHODS: Sensor to infant first (STIF) and then to oximeter was compared with sensor to oximeter first (STOF) and then to infant in $28 weeks gestations. The primary outcome was time from completion of sensor application to reliable signal, defin...

متن کامل

Pulse Oximeter Sensor Application During Neonatal Resuscitation: A Randomized Controlled Trial

OBJECTIVE: This study was done to compare 2 techniques of pulse oximeter sensor application during neonatal resuscitation for faster signal detection. METHODS: Sensor to infant first (STIF) and then to oximeter was compared with sensor to oximeter first (STOF) and then to infant in $28 weeks gestations. The primary outcome was time from completion of sensor application to reliable signal, defin...

متن کامل

A pilot randomized controlled trial of EKG for neonatal resuscitation

BACKGROUND The seventh edition of the American Academy of Pediatrics Neonatal Resuscitation Program recommends the use of a cardiac monitor in infants that need resuscitation. Previous trials have shown that EKG heart rate is available before pulse rate from a pulse oximeter. To date no trial has looked at how the availability of electrocardiogram (EKG) affects clinical interventions in the del...

متن کامل

Obtaining pulse oximetry data in neonates: a randomised crossover study of sensor application techniques.

Pulse oximetry may be useful during neonatal resuscitation. A randomised crossover study was performed to determine the most efficient method of applying the sensor. Applying it to the infant before connecting to the oximeter resulted in quickest acquisition of accurate heart rate. This technique should be preferred during resuscitation.

متن کامل

Use of a Combined SpO2/PtcCO2 Sensor in the Delivery Room

Arterial oxygen saturation (SaO(2)) and partial arterial pressure of carbon dioxide (PaCO(2)) are important respiratory parameters in critically ill neonates. A sensor combining a pulse oximeter with the Stow-Severinghaus electrode, required for the measurement of peripheral oxygen saturation (SpO(2)) and transcutaneous partial pressure of carbon dioxide (PtcCO(2)), respectively, has been recen...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Pediatrics

دوره 133 3  شماره 

صفحات  -

تاریخ انتشار 2014