Automatic measurement of blood pressure: evaluation of the Copal UA-231 automatic sphygmomanometer.
نویسندگان
چکیده
The suitability of an automatic sphygmomanometer for epidemiological work was evaluated in two studies. In the first, blood pressure measurements were made alternately on the automatic machine and on a Hawksley random zero instrument on 14 volunteers on each of four consecutive days by a single medical observer. No significant difference was found between the mean diastolic pressure obtained by each instrument but the mean systolic pressure was 5 mm higher on average on the automatic machine. In the second study, the effect of cuff position for the automatic instrument was examined. Differences between recommended and errant cuff positions of up to 4.5 mmHg were recorded but these did not reach statistical significance. These studies indicate that if used carefully this automatic instrument is suitable for survey and experimental use.
منابع مشابه
Field evaluation of the Copal UA-231 automatic sphygmomanometer.
Automated sphygmomanometers should be evaluated in field studies before being recommended for epidemiological use. In this study an automated sphygmomanometer, the Copal UA-231, was evaluated on the basis of duplicate blood pressure measurements, one with this machine and one with a Hawksley random zero machine, taken on 1536 participants in a health survey. The Copal-random zero differences ha...
متن کاملAgreement between the Takeda UA-731 automatic blood pressure measuring device and the manual mercury sphygmomanometer: an assessment under field conditions in Newcastle upon Tyne, UK.
STUDY OBJECTIVE To assess agreement between two Takeda UA-731 automatic blood pressure measuring devices (referred to as machines A and B) and two manual mercury sphygmomanometers. DESIGN A 'Y' connector attached each Takeda UA-731 to a manual mercury sphygmomanometer. Simultaneous measurements were made on adult subjects. SETTING A population based cardiovascular disease survey in Newcastl...
متن کاملMeasuring blood pressure at the wrist: more comfortable for patients and more convenient for doctors?
OBJECTIVES To compare the agreement between conventional measurement of blood pressure and measurements obtained using two automated devices; and to compare how comfortable each of the three methods of measurement were for patients. METHODS Blood pressure measurements and patient comfort scores were recorded using three different devices in 125-surgery and 40-community patients. The devices u...
متن کاملImproving the Design of Blood Pressure and Blood Saturation Monitors
A blood pressure monitor or sphygmomanometer can be either manual or automatic, employing respectively either the auscultatory method or the oscillometric method. The manual version of the sphygmomanometer involves an inflatable cuff with a stethoscope adopted to detect the sounds generated by the arterial walls to measure blood pressure in an artery. An automatic sphygmomanometer can be effect...
متن کاملغربالگری افزایش فشارخون در رانندگان کامیون
Introduction: Hypertension is a major risk factor for coronary heart disease. Truck drivers are more exposed to hypertension due to sedentary life style and consequent obesity, improper food, job stress and high prevalence of cigarette smoking. Materials and Methods: In this study, after completing a questionnaire by interview, the blood pressures of 122 members of ::::union:::: of truck driver...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of epidemiology and community health
دوره 39 3 شماره
صفحات -
تاریخ انتشار 1985