Transcutaneous bilirubinometry in preterm infants receiving phototherapy.

نویسندگان

  • Shakuntala Nanjundaswamy
  • Anna Petrova
  • Rajeev Mehta
  • Thomas Hegyi
چکیده

The accuracy of transcutaneous bilirubin (TcB) measurements obtained with a Bilicheck device in 70 preterm neonates before phototherapy and during phototherapy from the exposed and unexposed skin of the forehead (under the eye patch) was evaluated by comparing the TcB with the total serum bilirubin (TSB) that was performed within +/- 30 minutes of the TcB measurements. The correlation and proportion of similar, overestimated, and underestimated paired TcB/TSB measurements was calculated. A TSB point estimate of +/- 1.5 mg/dL was used for comparison of differences between the TcB and TSB. There was weaker correlation between TcB and TSB when the TcB was obtained from the exposed skin (r = 0.70; p > 0.05) as compared with an unexposed area (r = 0.77; p < 0.04) or before initiation of phototherapy (r = 0.86; p < 0.01). The TcB measurements from the area unexposed to phototherapy were predominantly (97.2%) in the similar (57%) or overestimation categories (40.2%). A significantly lower proportion of the paired TcB/TSB measurements obtained from exposed area were similar (42%) or overestimated (27%; p < 0.001). More importantly, in 31% of the measurements, the TSB levels were underestimated by the TcB. We conclude that the Bilicheck can be used for the evaluation of bilirubin levels in preterm neonates receiving phototherapy by using the unexposed skin of the forehead (under the eye patch) for TcB measurements; this may significantly reduce the number of blood draws.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Transcutaneous bilirubinometry in preterm neonates.

This prospective study was conducted to evaluate the accuracy of transcutaneous bilirubinometry in preterm newborns less than 32 weeks of gestation. Serum bilirubin values measured by direct spectrophotometry were considered as standard, the range was 2.2-12.5 mg/dL. 32 jaundiced infants of less than 32 weeks of gestation without phototherapy, including 10 ELBW neonates, were enrolled. Close co...

متن کامل

Transcutaneous Bilirubinometry in Preterm and Term Newborn Infants before and during Phototherapy.

INTRODUCTION To evaluate the accuracy of transcutaneous bilirubin measurement in a large population of newborn infants, before and during the phototherapy. PATIENTS AND METHODS A single Bilicheck instrument was used for transcutaneous measurements. A photo-opaque patch was positioned over the measurement site prior to starting phototherapy. Transcutaneous bilirubinometry was conducted on an u...

متن کامل

Diagnostic Usefulness of Transcutaneous Bilirubinometry in Very Preterm Newborns

BACKGROUND This study was performed to find out whether transcutaneous bilirubinometry could be a valid screening method for hyperbilirubinemia in preterm infants, especially for those who needed mechanical ventilation. METHODS We evaluated 63 preterm Iranian newborns who were managed in the neonatal intensive care unit of Shahidbeheshti University Hospital, Isfahan, Iran from April 2009 to A...

متن کامل

Evaluation of transcutaneous bilirubinometry in preterm infants of gestational age 32-34 weeks.

AIMS The aim of this prospective study was to evaluate the accuracy of transcutaneous bilirubinometry using the Minolta Air-Shields JM-103 device in preterm newborns of gestational age 32-34 weeks, and to identify the most appropriate measurement site. METHODS Transcutaneous bilirubin (TcB) measurements were performed over forehead, sternum and abdomen, if total serum bilirubin (TSB) had to b...

متن کامل

Transcutaneous bilirubin after phototherapy in term and preterm infants.

OBJECTIVES To compare transcutaneous bilirubin (TcB) readings with total serum bilirubin (TSB) after phototherapy, estimating the range of TcB where confirmation through blood sampling can be avoided. METHODS Preterm and term neonates receiving in-hospital phototherapy underwent TcB measurements (device JM-103, TcB) alongside routine TSB before and after treatment. We calculated time-dependen...

متن کامل

ذخیره در منابع من