Evaluation of four gentamicin and tobramycin assay procedures for clinical laboratories.

نویسندگان

  • F G Witebsky
  • C A Sliva
  • S T Selepak
  • M E Ruddel
  • J D MacLowry
  • E E Johnson
  • R J Elin
چکیده

Accuracy, precision, and clinical laboratory utility of the TDX (Abbott Laboratories), Auto-ICS (Beckman Instruments, Inc.), COBAS-Bio (Roche Analytical Instruments, Inc.) with reagent kits (Syva), and EMIT (Syva) for gentamicin and tobramycin serum assay were assessed. TDX, COBAS-Bio, and EMIT analytical systems showed a proportional bias of less than 10% for recovery studies and a coefficient of variation less than 5% for within-run precision. The results of the recovery studies with the Auto-ICS showed a proportional bias of 25% with gentamicin and 16% with tobramycin. The within-run precision expressed as the coefficient of variation for the Auto-ICS was 6.7% for gentamicin and 8.6% for tobramycin. In comparisons involving gentamicin- and tobramycin-containing patient samples, the results with the TDX analytical system showed the best agreement with the COBAS-Bio. For the determination of these two antibiotics, the TDX analytical system provided the best overall accuracy and precision.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Analysis of neomycin, kanamycin, tobramycin and amikacin resistance mechanisms in gentamicin-resistant isolates of Enterobacteriaceae.

Twenty-four gentamicin-resistant isolates of Enterobacteriaceae, obtained from the clinical laboratories of three health centres in Nablus, Palestine, were tested for susceptibility to neomycin, kanamycin, tobramycin and amikacin. Resistance rates were 29.2% for neomycin, 58.3% for kanamycin, 45.8% for tobramycin and 8.3% for amikacin. Fourteen (58.3%) isolates were noted to be multiresistant, ...

متن کامل

Evaluation of five gentamicin assay procedures for clinical microbiology laboratories.

Five gentamicin assay procedures (a bioassay, an enzyme immunoassay, a latex agglutination inhibition test, a fluorescence immunoassay, and a radioimmunoassay) were evaluated to determine which was optimal for our laboratory. The evaluation was based on recovery and precision studies and results of analyses of patient samples, as well as technical assay performance factors. The latex agglutinat...

متن کامل

Serum standards for the bioassay of aminoglycosides in cerebrospinal fluid.

Four diluents were compared as reference standards for the assay of gentamicin in cerebrospinal fluid (CSF): human CSF, human serum, distilled water, and 150 mmol NaCl/4.5 mmol CaCl2. Standards prepared in pooled human serum were the best alternative to CSF for the assay of gentamicin and were also useful for the assay of tobramycin, netilmicin, amikacin, and sisomicin. The pH (6.0-9.8) of CSF ...

متن کامل

Evaluation of four assay methods for determination of tobramycin in human serum.

Four assay procedures for tobramycin in serum--enzyme immunoassay (I), substrate-labeled fluorescent immunoassay (II), radioimmunoassay (III), and bioassay (IV)--were compared and evaluated by replicate and analytical recovery studies. I and II were about 50% more precise than III and IV. II was substantially more nearly accurate than the other methods and also gave the best reproducibility (co...

متن کامل

Measurement of three aminoglycosiee antibiotics with a single radioimmunoassay system.

A versatile radioimmunoassay procedure has been developed to specifically assay gentamicin, tobramycin, or amikacin. Iodinated derivatives of these three antibiotics were prepared by a modified Bolton and Hunter procedure. A single reagent system composed of a pool of three very specific antisera was used in the procedure. The antisera combination did not alter the specificity of the procedure....

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of clinical microbiology

دوره 18 4  شماره 

صفحات  -

تاریخ انتشار 1983