Establishment of Age- and Gender-Specific Reference Ranges for 36 Routine and 57 Cell Population Data Items in a New Automated Blood Cell Analyzer, Sysmex XN-2000.

نویسندگان

  • Sang Hyuk Park
  • Chan Jeoung Park
  • Bo Ra Lee
  • Mi Jeong Kim
  • Min Young Han
  • Young Uk Cho
  • Seongsoo Jang
چکیده

We established age- and gender-specific reference ranges for the 36 routine complete blood cell (CBC) and 57 cell population data (CPD) items in the Sysmex XN-2000 (Sysmex, Japan). In total, 280 peripheral blood samples were obtained from an equal number of healthy adults. Values for 36 routine items and 57 CPD items were obtained for each sample, and the results were categorized into six subgroups (N>39 in each subgroup) according to patient age (20-40, 41-60, and >60 yr) and gender (male and female), and compared with respect to age and gender differences. The majority of data items (22 of 36 routine CBC items and 44 of 57 CPD items) exhibited significant differences (P≤0.05) in their results with respect to age or gender, and several red cell-, lymphocyte-, and platelet-related data tended to decrease in women or older adults. These results provide a basis for establishing age- and gender-specific reference ranges for routine and CPD items in Sysmex XN-2000. Furthermore, these reference ranges could be used to determine clinical significance for new items of Sysmex XN-2000 in further studies.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The New Sysmex XN-2000 Automated Blood Cell Analyzer More Accurately Measures the Absolute Number and the Proportion of Hematopoietic Stem and Progenitor Cells Than XE-2100 When Compared to Flow Cytometric Enumeration of CD34+ Cells

The gold standard to count circulating hematopoietic stem and progenitor cells (HPCs) is to enumerate CD34 cells by using flow cytometry [1, 2]. XE-2100 (Sysmex, Kobe, Japan), a currently used hematology analyzer, can be used to quantify HPCs, generating counts that reportedly correlate with those obtained using flow cytometric analysis of CD34 cells; however, XE-2100 HPC counts were 1.86-fold ...

متن کامل

Hematological Reference Values for Healthy Males in the Central Part of Iran

  Background and Objectives: Laboratory reference values are an important tool for clinical management of patients. Reference values being used in most laboratories in Iran have been provided from the established reference values from developed countries. However, several variables can affect on these laboratory parameters. Therefore, this study was carried out to establish the reference value...

متن کامل

Reference interval for immature platelet fraction on Sysmex XN haematology analyser in adult population

Introduction The Sysmex XN-series haematology analyser has newly adopted a fluorescent channel to measure immature platelet fraction (IPF). To promote the clinical utility of this promising parameter, establishing a reliable reference interval is mandatory. According to previous studies, IPF values may be affected by the employed analyser and the ethnic background of the individual, but no diff...

متن کامل

Validity of Selected WBC Differentiation Flags in Sysmex XT-1800i

Background: Automatic Cell Counter devises make the CBC differential very easy and delivering the results in few second. However, the problem with this device is facing a flag requires a time-consuming microscopic review of the specimen which causes unacceptable wait times for patient as well as costs for laboratories. In this study, we calculated the validity of WBC d...

متن کامل

Reference Values for Peripheral Blood Lymphocyte Subsets in a Healthy Korean Population

Flow cytometric immunophenotyping of peripheral blood lymphocyte subsets is a powerful tool for evaluating cellular immunity and monitoring immune-mediated diseases. The numbers and proportions of blood lymphocyte subsets are influenced by factors such as gender, age, ethnicity, and lifestyle. This study aimed to establish reference ranges for peripheral blood lymphocyte subsets in a healthy Ko...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Annals of laboratory medicine

دوره 36 3  شماره 

صفحات  -

تاریخ انتشار 2016