Percutaneous extraction of a coiled, 20-year-old lead in a patient with cardiac resynchronization therapy.

نویسندگان

  • Barbara Małecka
  • Andrzej Kutarski
  • Jacek Bednarek
  • Andrzej Zabek
  • Jacek Lelakowski
چکیده

A 61-year-old patient with a 20-year history of permanent pacemaker implantation and half--a-year cardiac resynchronization therapy using a left ventricular lead placed via surgical approach was admitted for extraction of an old coiled right ventricular lead, which triggered ventricular arrhythmia and created a risk of pulmonary embolism. The lead was extracted via the left femoral vein in two stages: untying a loop on the lead using a pig-tail catheter and Dotter basket followed by traction and dissection of adhesions using a Byrd dilator sheath. Dissection of the old lead from the active right ventricular one posed special technical problems.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Successful transvenous lead extraction after a failed open surgical attempt.

An 81-year-old man was referred with an infected cardiac resynchronization therapy defibrillator (CRT-D) system. His initial dualchamber implantable cardioverter-defibrillator (ICD) had been implanted 8 years earlier and upgraded to a CRT-D system 2 years before referral. Epicardial atrial and left ventricular (LV) leads were subsequently added after the endocardial LV lead lost capture. The en...

متن کامل

Successful percutaneous repositioning of a dislodged atrial pacemaker lead with a deflectable catheter.

A 67-year-old male patient with ischemic dilated cardiomyopathy underwent implantation of an implantable cardioverter-defibrillator with cardiac resynchronization therapy about one month ago. The implantation process was successful and the patient was discharged in good condition. At the routine first month follow-up, pacemaker analysis revealed that there was no atrial pacing and sensing, alth...

متن کامل

Thoracoscopic Left Ventricular Lead Implantation in Cardiac Resynchronization Therapy

Cardiac resynchronization therapy is known to reduce morbidity and mortality in patients with advanced heart failure as a result of dyssynchrony and systolic dysfunction of the left ventricle. Placement of the left ventricular (LV) lead via the coronary sinus can be difficult. When LV lead implantation is difficult, a video-assisted epicardial approach can be a good alternative. Although there ...

متن کامل

Simple and effective solution for diaphragm pacing by that uses the transvenous femoral approach in a patient treated with cardiac resynchronization-defibrillator therapy: The wrapping method

The use of cardiac resynchronization therapy (CRT) has become widespread in patients with heart failure who meet the criteria for implantation. Coronary sinus lead positioning is important to provide the intended biventricular stimulation. Diaphragm pacing is a lead-related complication which occurs secondary to phrenic nerve stimulation. We present the case of a 58-year-old male patient who ha...

متن کامل

Pacemaker-mediated tachycardia in an unconventional resynchronization device.

A 67-year-old male patient with heart failure and permanent atrial fibrillation underwent atrio-ventricular nodal ablation and a biventricular device implantation. Instead of a standard cardiac resynchronization therapy device, a DDD pacemaker was implanted with the left ventricular lead in the atrial channel. After the implant, a pacemaker-mediated tachycardia caused by T wave oversensing by t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Cardiology journal

دوره 15 4  شماره 

صفحات  -

تاریخ انتشار 2008