Angle Kappa Measurements: Normal Values in Healthy Iranian Population Obtained With the Orbscan II
نویسندگان
چکیده
BACKGROUND The angle kappa is important in proper centration of corneal ablation in keratorefractive surgery. Orbscan II device is widely used preoperatively in photoablation surgeries and can be used to measure the angle kappa. OBJECTIVES This study aimed to determine the mean angle kappa and its intercepts in healthy young Iranian adults. PATIENTS AND METHODS In this cross-sectional study, orthotropic patients (age range, 18-35 years) who were referred to the Khatam Eye Hospital (Mashhad, Iran) were included. Exclusion criteria were as follows: history of any eye deviation or strabismus with or without orthoptic or surgical treatment; any intraocular, corneal, or keratorefractive surgery; contact lens use; any corneal anomaly; any ophthalmic or systemic drug consumption; and hyperopic spherical refraction > + 3.00 diopters (D), spherical refraction > -5.00 D, or cylindrical refraction > 2.00 D. All of the parameters were measured by the same operator through an Orbscan II device. RESULTS A total of 977 healthy participants who aged 18 to 45 years were included consecutively. The study population consisted of 614 females and 363 males. The average angle kappa was 5.00º ± 1.36º at 240.21º ± 97.17º in males and 4.97º ± 1.30º at 244.22º ± 94.39º in females (P = 0.63). The average horizontal (x-axis) angle kappa was -0.02º ± 0.49º, with a mean of -0.02º ± 0.50º in males and -0.02º ± 0.49º in females (P = 0.93). The average vertical (y-axis) angle kappa was -0.09º ± 0.32º, with a mean of -0.09º ± 0.33º in males and -0.09º ± 0.32º in females (P = 0.74). CONCLUSIONS By using the normal angle kappa determined in this study, pseudodeviations can be identified more precisely in those who might undergo keratorefractive surgery.
منابع مشابه
Measurement of angle kappa with synoptophore and Orbscan II in a normal population.
PURPOSE To obtain normative values of angle kappa in a normal population by synoptophore and Orbscan II and to compare the reliability of these devices. METHODS Three hundred consecutive healthy individuals were enrolled in the study. A complete orthoptic and ophthalmologic examination was performed. Synoptophore and Orbscan II corneal topography were used to measure angle kappa. To evaluate ...
متن کاملWhite-to-white corneal diameter: normal values in healthy Iranian population obtained with the Orbscan II.
AIM To determine the normative values of white-to-white corneal diameter with Orbscan II Topography System and to compare right and left eyes data in the normal young population. METHODS A total of 1001 healthy participants aged 18-45y participated in this observational cross-sectional study. The study population consisted of 616 female and 385 male subjects. The corneal diameter was measured...
متن کاملMeasurement of Angle Kappa Using Ultrasound Biomicroscopy and Corneal Topography
PURPOSE To introduce a new convenient and accurate method to measure the angle kappa using ultrasound biomicroscopy (UBM) and corneal topography. METHODS Data from 42 eyes (13 males and 29 females) were analyzed in this study. The angle kappa was measured using Orbscan II and calculated with UBM and corneal topography. The angle kappa of the dominant eye was compared with measurements by Orbs...
متن کاملPhase Angle Measurement in Healthy Human Subjects through Bio-Impedance Analysis
Objective(s)Bioelectrical impedance is the measure of impedance of the body. Impedance consists of electric resistance and reactance. Phase angle (PA) is the tan value of the ratio of reactance versus electric resistance. PA depends on cell membrane integrity and on body cell mass. There exists a correlation between PA values and body cell mass.The objective of this study was to compare the PA ...
متن کاملTopographic paracentral corneal thickness with pentacam and orbscan: effect of acoustic factor.
PURPOSE To evaluate the effect of an acoustic factor (AF) on the comparison of central corneal thickness (CCT) and peripheral corneal thickness (PCT) measurements with Orbscan II and Pentacam. METHODS The CCT and PCT at 1, 2, and 3 radial distances from the corneal apex were measured using Orbscan II and Pentacam in 22 right eyes of 22 healthy adults (7 men, 15 women). Three measures were obt...
متن کامل