A motion phantom study on helical tomotherapy: the dosimetric impacts of delivery technique and motion.
نویسندگان
چکیده
Helical tomotherapy (HT) can potentially be used for lung cancer treatment including stereotactic radiosurgery because of its advanced image guidance and its ability to deliver highly conformal dose distributions. However, previous theoretical and simulation studies reported that the effect of respiratory motion on statically planned tomotherapy treatments may cause substantial differences between the calculated and actual delivered radiation isodose distribution, particularly when the treatment is hypofractionated. In order to determine the dosimetric effects of motion upon actual HT treatment delivery, phantom film dosimetry measurements were performed under static and moving conditions using a clinical HT treatment unit. The motion phantom system was constructed using a programmable motor, a base, a moving platform and a life size lung heterogeneity phantom with wood inserts representing lung tissue with a 3.0 cm diameter spherical tumour density equivalent insert. In order to determine the effects of different motion and tomotherapy delivery parameters, treatment plans were created using jaw sizes of 1.04 cm and 2.47 cm, with incremental gantry rotation periods between the minimum allowed (10 s) and the maximum allowed (60 s). The couch speed varied from 0.009 cm s(-1) to 0.049 cm s(-1), and delivered to a phantom under static and dynamic conditions with peak-to-peak motion amplitudes of 1.2 cm and 2 cm and periods of 3 and 5 s to simulate human respiratory motion of lung tumours. A cylindrical clinical target volume (CTV) was contoured to tightly enclose the tumour insert. 2.0 Gy was prescribed to 95% of the CTV. Two-dimensional dose was measured by a Kodak EDR2 film. Dynamic phantom doses were then quantitatively compared to static phantom doses in terms of axial dose profiles, cumulative dose volume histograms (DVH), percentage of CTV receiving the prescription dose and the minimum dose received by 95% of the CTV. The larger motion amplitude resulted in more under-dosing at the ends of the CTV in the axis of motion, and this effect was greater for the smaller jaw size plans. Due to the size of the penumbra, the 2.47 cm jaw plans provide adequate coverage for smaller amplitudes of motion, +/-0.6 cm in our experiment, without adding any additional margin in the axis of motion to the treatment volume. The periodic heterogeneous patterns described by previous studies were not observed from the single fraction of the phantom measurement. Besides the jaw sizes, CTV dose coverage is not significantly dependent on machine and phantom motion periods. The lack of adverse synchronization patterns from both results validate that HT is a safe technique for treating moving target and hypofractionation.
منابع مشابه
A comparison of the dosimetric effects of intrafraction motion on step‐and‐shoot, compensator, and helical tomotherapy‐based IMRT
Intrafraction motion during intensity-modulated radiation therapy can cause differences between the planned and delivered patient dose. The magnitude of these differences is dependent on a number of variables, including the treatment modality. This study was designed to compare the relative susceptibility of plans generated with three different treatment modalities to intrafraction motion. The ...
متن کاملA feasibility study of dynamic verification for tumor target delineation and dose delivery using a six degrees of freedom motion phantom
Background: The dynamic phantom is one of the best tools to study the impact of motion on tumor target delineation and absorbed dose verification during dose delivery. Materials and Methods: this study, a 6-DOF (degrees of freedom) phantom was designed following the stacked serial kinematics and assembled by six commercial motion stages to generate 6-DOF motion, which were RotX (pitch, around X...
متن کاملInvestigation of tumor motion influence on applied dose distribution in conventional proton therapy vs. IMPT a 4D Monte Carlo simulation study
Background: in radiation treatment of moving targets located in thorax region of patient body, the delivered dose does not match with the planned treatment, resulting in some over and under dosage in the tumor volume, as a function of motion magnitude and frequency. Several efforts have been done to investigate the target motion effects on dose distribution in the target and surrounding normal ...
متن کاملA study of longitudinal tumor motion in helical tomotherapy using a cylindrical phantom
Tumor motion during radiation treatment on a helical tomotherapy unit may create problems due to interplay with motion of the multileaf collimator, gantry rotation, and patient couch translation through the gantry. This study evaluated this interplay effect for typical clinical parameters using a cylindrical phantom consisting of 1386 diode detectors placed on a respiratory motion platform. All...
متن کاملDosimetric verification and quality assurance of running‐start‐stop (RSS) delivery in tomotherapy
The purpose of this study was to evaluate the dosimetric profiles and delivery accuracy of running-start-stop (RSS) delivery in tomotherapy and to present initial quality assurance (QA) results on the accuracy of the dynamic jaw motion, dosimetric penumbrae of the RSS dynamic jaw and the static jaw were measured by radiographic films. Delivery accuracy of the RSS was evaluated by gamma analysis...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Physics in medicine and biology
دوره 52 1 شماره
صفحات -
تاریخ انتشار 2007