Minimal inspiratory flow from dry powder inhalers according to a biphasic model of pressure vs. flow relationship.
نویسندگان
چکیده
Inhalation therapy using the dry powder inhaler (DPI) is now the first choice for obstructive pulmonary diseases. We previously measured relationships between inspiratory pressure (PI) and flow rate of almost all of the DPIs available in Japan, and described an importance of inspiratory efforts. In the present study, we further analyzed the data obtained in the previous study. Although there were linear relationships between PI and flow2, the slope became steeper when PI was less than a certain value (critical PI, existed between 15-20 cmH2O). When PI was less than critical PI, linear rather than parabolic regression between PI and flow yielded better fits (r > 0.90, p < 0.001). Inspiratory flows at the critical PI were 53.9 (Diskus), 65.8 (Diskhaler), 45.9 (Turbuhaler for Pulmincort), 48.6 (Turbuhaler for Symbicort) and 38.0 l/min (Twisthaler). These findings suggested that flow through the DPI becomes laminar rather than turbulent flow in the range below critical PIs. We suggest that patients should inhale from the DPIs with inspiratory pressure higher than critical PI.
منابع مشابه
Peak inspiratory flow rate and slope of the inhalation profiles in dry powder inhalers.
Aerosol delivery depends on device design and inhalation technique. In vitro device evaluations have shown that the emitted dose and fine particle mass of dry powder inhalers (DPIs) increase at high peak inspiratory flow rates (PIFR). Since the PIFR is mostly achieved after the release of the powder, slope of the pressure profile is also described as an important determinant. The aim of the pre...
متن کاملInspiratory flow rates at different levels of resistance in elderly COPD patients.
Dry powder inhalers (DPIs) are increasingly replacing metered dose inhalers in elderly chronic obstructive pulmonary disease (COPD) patients. However, most DPIs are dependent on inspiratory flow, which is compromised by the ageing process itself. Using the in-check dial method, the present study compared peak inspiratory flow (PIF) rates in 26 elderly COPD patients and 14 matched control subjec...
متن کاملInspiratory flows through dry powder inhaler in chronic obstructive pulmonary disease: age and gender rather than severity matters
BACKGROUND Dry powder inhalers (DPIs) are inspiratory flow driven and hence flow dependent. Most patients with chronic obstructive pulmonary disease (COPD) are elderly and have poor lung function. The factors affecting their inspiratory flows through inhalers are unclear. OBJECTIVE To study peak inspiratory flows (PIFs) and their determinants through a DPI in COPD patients of varying age and ...
متن کاملPeak inspiratory flow through Turbuhaler in acute asthma.
Efficient use of dry powder inhalers, such as Turbuhaler, is dependent on the generation of adequate inspiratory flow. It is not clear whether patients with acute asthma are able to generate adequate flow. Peak inspiratory flow (PIF) was measured through an empty Turbuhaler, and without this device, in 99 adults presenting to hospital with acute exacerbations of asthma. Where possible, patients...
متن کاملScintigraphic comparison of budesonide deposition from two dry powder inhalers.
Chlorofluorocarbons (CFCs), used in metered dose inhalers (MDIs), have been identified as being deleterious to the environment leading to a ban on their production. Dry powder inhalers (DPIs) are a widely used alternative to MDIs. One disadvantage of DPIs is that in vivo lung deposition can be influenced by the patient's inspiratory flow rate. The ASTA Medica multi-dose dry powder inhaler (AM-M...