Evaluation of fluence‐based dose delivery incorporating the spatial variation of dosimetric leaf gap (DLG)

نویسندگان

  • Lalith K. Kumaraswamy
  • Zhengzheng Xu
  • Daniel W. Bailey
  • Jonathan D. Schmitt
  • Matthew B. Podgorsak
چکیده

The Eclipse treatment planning system uses a single dosimetric leaf gap (DLG) value to retract all multileaf collimator leaf positions during dose calculation to model the rounded leaf ends. This study evaluates the dosimetric impact of the 2D variation of DLG on clinical treatment plans based on their degree of fluence modulation. In-house software was developed to retrospectively apply the 2D variation of DLG to 61 clinically treated VMAT plans, as well as to several test plans. The level of modulation of the VMAT cases were determined by calculating their modulation complexity score (MCS). Dose measurements were done using the MapCHECK device at a depth of 5.0 cm for plans with and without the 2D DLG correction. Measurements were compared against predicted dose planes from the TPS using absolute 3%/3 mm and 2%/2 mm gamma criteria for test plans and for VMAT cases, respectively. The gamma pass rate for the 2 mm, 4 mm, and 6 mm sweep test plans increased by 23.2%, 28.7%, and 26.0%, respectively, when the measurements were corrected with 2D variation of DLG. The clinical anal VMAT cases, which had very high MLC modulation, showed the most improvement. The majority of the improvement occurred for doses created by the 1.0 cm width leaves for both the test plans and the VMAT cases. The gamma pass rates for the highly modulated head and neck (H&N) cases, moderately modulated prostate and esophageal cases, and minimally modulated brain cases improved only slightly when corrected with 2D variation of DLG. This is because these cases did not employ the 1.0 cm width leaves for dose calculation and delivery. These data suggest that, at the very least, the TPS plans with highly modulated fluences created by the 1.0 cm fields require 2D DLG correction. Incorporating the 2D variation of DLG for the highly modulated clinical treatment plans improves their planar dose gamma pass rates, especially for fields employing the outer 1.0 cm width MLC leaves. This is because there are differences in DLG between the true DLG exhibited by the 1.0 cm width outer leaves and the constant DLG value modeled by the TPS for dose calculation.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Spatial variation of Dosimetric Leaf Gap and the impact in Dose delivery Our work evaluates the variation of Dosimetric Leaf Gap (DLG) along the travel path of each

Our work evaluates the variation of Dosimetric Leaf Gap (DLG) along the travel path of each Varian Millennium 120-MLC leaf pairs. Traditionally, the Treatment Planning Systems (TPS) utilize a single value of DLG measured under the middle leaf-pair. The TPS models the shape of the leaf tip as straight as opposed to rounded edge. To overcome this limitation, the TPS retracts each leaf tip by half...

متن کامل

On using the dosimetric leaf gap to model the rounded leaf ends in VMAT/RapidArc plans

Partial transmission through rounded leaf ends of Varian multileaf collimators (MLC) is accounted for with a parameter called the dosimetric leaf gap (DLG). Verification of the value of the DLG is needed when the dose delivery is accompanied by gantry rotation in VMAT plans. We compared the doses measured with GAFCHROMIC film and an ionization chamber to treatment planning system (TPS) calculat...

متن کامل

The dosimetric impact of control point spacing for sliding gap MLC fields

Dynamic sliding gap multileaf collimator (MLC) fields are used to model MLC properties within the treatment planning system (TPS) for dynamic treatments. One of the key MLC properties in the Eclipse TPS is the dosimetric leaf gap (DLG) and precise determination of this parameter is paramount to ensuring accurate dose delivery. In this investigation, we report on how the spacing between control ...

متن کامل

Optimization of the dosimetric leaf gap for use in planning VMAT treatments of spine SABR cases

The dosimetric leaf gap (DLG) is a beam configuration parameter used in the Varian Eclipse treatment planning system, to model the effects of rounded MLC leaf ends. Measuring the DLG using the conventional sliding-slit technique has been shown to be produce questionable results for some volumetric modulated arc therapy (VMAT) treatments. This study therefore investigated the use of radiochromic...

متن کامل

Tuning of AcurosXB source size setting for small intracranial targets

This study details a method to evaluate the source size selection for small field intracranial stereotactic radiosurgery (SRS) deliveries in Eclipse treatment planning system (TPS) for AcurosXB dose calculation algorithm. Our method uses end-to-end dosimetric data to evaluate a total of five source size selections (0.50 mm, 0.75 mm, 1.00 mm, 1.25 mm, and 1.50 mm). The dosimetric leaf gap (DLG) ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2016