Performance of the Statstrip Glucose Meter in Inpatient Management of Diabetes Mellitus
نویسندگان
چکیده
The use of hospital point-of-care glucose meters in inpatient care is widely established despite unsolved issues regarding accuracy. The aim of this study was to validate the performance of the StatStrip Glucose Hospital Meter (Nova Biomedical, Waltham, USA) in performing glycemic profile in hospitalized diabetic patients undergoing intensified insulin therapy. We investigated total imprecision, inaccuracy, and analytical range limits by measuring glucose in 481 plasma samples of diabetic patients. Total imprecision was 2.4% and linearity ranged from 1.3 to 31.9 mmol/L. The results correlated well (r=0.9868) with the laboratory routine procedure with a determined bias of 2.1%. Total error of the method was 4.5%, which was within designated criteria for laboratory glucose measurement. Diagnostic testing revealed a 100% sensitivity and 100% specificity in detection of hypoglycemic episodes (n=16), respectively. StatStrip Glucose Hospital Meter is a simple, accurate and reliable tool for assessing glycemia levels in inpatient management of diabetes mellitus.
منابع مشابه
Validation of Point-of-Care Glucose Testing for Diagnosis of Type 2 Diabetes
Point-of-care (POC) glucose technology is currently considered to be insufficiently accurate for the diagnosis of diabetes. The objective of this study was to investigate the diagnostic accuracy of an innovative, interference-resistant POC glucose meter (StatStrip glucose hospital meter, Nova Biomedical, USA) in subjects with a previous history of dysglycaemia, undergoing a 75 g diagnostic oral...
متن کاملPerformance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus
OBJECTIVE To evaluate the performance of the FreeStyle Libre Flash continuous glucose monitoring (FSL-CGM) system against established central laboratory methods. RESEARCH DESIGN AND METHODS 20 subjects (8 type 1 diabetes mellitus, 12 type 2 diabetes mellitus) were analyzed. FSL-CGM sensor measurements (inserted in arm and abdomen) were compared with capillary blood glucose results analyzed wi...
متن کاملAnalytical Performance and Quality Control of a Glucose Monitor System
Background and Objective: The reliability and validity of monitors for self-monitoring of blood glucose are debated. We evaluated the analytical performance of Accu-check Active (Boehringer Mannheim, Roche) which is one of the most commonly used monitors in Iran. Material and Methods: We compared the monitor readings with the reference values by percentage of values within...
متن کاملAccuracy of Capillary and Arterial Whole Blood Glucose Measurements Using a Glucose Meter in Patients under General Anesthesia in the Operating Room.
BACKGROUND The aim of this study was to evaluate the use of a glucose meter with surgical patients under general anesthesia in the operating room. METHODS Glucose measurements were performed intraoperatively on 368 paired capillary and arterial whole blood samples using a Nova StatStrip (Nova Biomedical, USA) glucose meter and compared with 368 reference arterial whole blood glucose measureme...
متن کاملEvaluation of the Nova Biomedical Statstrip glucose meter
Aims: The Nova StatStrip blood glucose monitoring system was evaluated for precision, accuracy, and interferences from haematocrit and maltose. It was also compared against three other meters and two reference methods. Methods: Heparinised whole blood samples were analysed on the meters. These results were compared with whole blood samples analysed on the Radiometer ABL835 for the interference ...