An X-ray flexure-based microgap rheometer (X-FMR) has been designed for combining rheology and in situ small-angle X-ray scattering from the vorticity plane. The gap distance can be varied continuously from 500 μm down to several μm, which provides the unique possibility to generate a strong confinement for many complex fluids. A singular advantage of this setup is the possibility to directly probe the vorticity direction of the flow field with a microfocus X-ray beam and to probe the structural response of the fluid to combined shear and confinement in the vorticity plane. The sliding-plate setup operates over a wide range of shear rates of γ = 10(-3)-10(3) s(-1) and strains in the range of 10(-4)-10(2). The flexure-based bearing maintains the plate parallelism within 10(-5) rad. The X-FMR requires very small sample volumes on the order of 10 μl. The applicability of the device is demonstrated here with limited examples of a nematic suspension of fd virus (rods), and a crystalline suspension containing sterically stabilized polystyrene-butylacrylate latex particles.
منابع مشابه
Shear-induced mesostructure in nanoplatelet-polymer networks
The shear response of a model polymer–clay gel is measured using small-angle neutron and light scattering, optical microscopy, and rheometry. As the flow disrupts the transient network that forms between clay and polymer, coupling between composition and stress leads to the formation of a macroscopic domain pattern, while the clay platelets orient with their surface normal parallel to the direc...
متن کاملShear-induced structure in polymer-clay nanocomposite solutions.
The equilibrium structure and shear response of model polymer-clay nanocomposite gels are measured using X-ray scattering, light scattering, optical microscopy, and rheometry. The suspensions form physical gels via the "bridging" of neighboring colloidal clay platelets by the polymer, with reversible adsorption of polymer segments onto the clay surface providing a short-range attractive force. ...
متن کاملApplication of small angle X-ray scattering (SAXS) for differentiation between normal and cancerous breast tissues
ABSTRACT Background: Coherent scattering leads to diffraction effects and especially constructive interferences. Theseinterferences carry some information about the molecular structure of the tissue. As breast cancer isthe most widespread cancer in women, this project evaluated the application of small angleX-ray scattering (SAXS) for differentiation between normal and cancerous breast tissues....
متن کاملIntroducing an Optimized Method for Obtaining X-ray Diffraction Patterns of Biological Tissues
Introduction Individual X-Ray diffraction patterns of biological tissues are obtained via interference of coherent scattering with their electrons. Many scientists have distinguished normal and cancerous breast tissue, bone density, and urinary stone types using the X-Ray diffraction patterns resulting from coherent scattering. The goal of this study was to introduce an optimized method for obt...
متن کاملBiaxial shear of confined colloidal hard spheres: the structure and rheology of the vorticity-aligned string phase.
Using a novel biaxial confocal rheoscope, we investigate the flow of the shear induced vorticity aligned string phase [X. Cheng et al., Proc. Natl. Acad. Sci. U. S. A., 2011, 109, 63], which has a highly anisotropic microstructure. Using biaxial shear protocols we show that we have excellent control of the string phase anisotropic morphology. We choose a shear protocol that drives the system in...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Review of scientific instruments
دوره 85 6 شماره
صفحات -
تاریخ انتشار 2014