Compression coil provides increased lead control in extraction procedures.
نویسندگان
چکیده
AIMS We investigated a new lead extraction tool (Compression Coil; One-Tie, Cook Medical) in an experimental traction force study. METHODS AND RESULTS On 13 pacemaker leads (Setrox JS53, Biotronik) traction force testing was performed under different configurations. The leads were assigned to three groups: (i) traction force testing without central locking stylet support (n = 5), (ii) traction force testing with the use of a locking stylet (Liberator, Cook Medical) and a proximal ligation suture (n = 4), (iii) traction force testing with the use of a locking stylet and a compression coil (n = 4). The following parameters were obtained for all groups: stress-strain curves, maximal forces, elastic modulus, post-testing lead length and lead elongation. In Groups 2 and 3 retraction of the locking stylet within the lead was measured [lead tip-locking stylet distance (LTLSD)]. Maximal forces for the three groups were: (i) 28.3 ± 0.3 N; (ii) 30.6 ± 3.0 N; (iii) 31.6 ± 2.9 N (1 vs. 2, P = 0.13; 1 vs. 3, P = 0.04; 2 vs. 3, P = 0.65). Elastic modulus was (i) 22.8 ± 0.1 MPa; (ii) 2830.8 ± 351.1 MPa; (iii) 2447.0 ± 510.5 MPa (1 vs. 2, P < 0.01; 1 vs. 3, P < 0.01; 2 vs. 3, P = 0.26). Mean LTLSD in Group 2 was 19.8 ± 3.2 cm and was 13.8 ± 1.7 cm in Group 3 (P = 0.02). The ratio of LTLSD/post-testing lead length was 0.37 ± 0.03 for Group 2 and 0.24 ± 0.03 for Group 3 (P < 0.01). CONCLUSION The application of a compression coil leads to an increased lead control expressed by less retraction of the locking stylet within the lead. This enables improved central support of extraction sheaths in the case of challenging extraction procedures.
منابع مشابه
Review of preconcentration and solid phase extraction for the determination of trace Lead
Spectrometric techniques for the analysis of trace lead have developed rapidly due to the increasing need for accurate measurements at extremely low levels of this element in diverse matrices. This review covers separation and preconcentration procedures, and considers the features of the application with several spectrometric techniques. The use of an appropriate sample handling technique is a...
متن کاملReview of preconcentration and solid phase extraction for the determination of trace Lead
Spectrometric techniques for the analysis of trace lead have developed rapidly due to the increasing need for accurate measurements at extremely low levels of this element in diverse matrices. This review covers separation and preconcentration procedures, and considers the features of the application with several spectrometric techniques. The use of an appropriate sample handling technique is a...
متن کاملTransvenous extraction of pacemaker and implantable cardioverter defibrillator leads using Evolution® mechanical dilator sheath: a single center confirmatory experience
OBJECTIVE In recent years there has been an increase in clinical situations requiring lead extraction procedures of implanted cardiac devices. In our clinic, extraction procedures are performed with Evolution® mechanical lead extraction system. In this manuscript we aimed to evaluate our lead extraction procedures. METHODS We retrospectively evaluated lead extraction procedures carried out on...
متن کاملSuperior vena cava defibrillator coils make transvenous lead extraction more challenging and riskier.
To the Editor: Studies have demonstrated equivalent defibrillation efficacy and all-cause mortality in patients with single and dual coil implantable cardioverter-defibrillator (ICD) leads (1,2). Despite this equivalency, the vast majority of implanted ICD leads are dual coil (3). The widespread use of dual coil ICD leads is relevant to lead management because of the challenges and risks of tra...
متن کاملSensitivity Analysis and Stray Capacitance of Helical Flux Compression Generator with Multi Layer Filamentary Conductor in Rectangular Cross-Section
This paper presents an approach to calculate the equivalent stray capacitance (SC) of n-turn of the helical flux compression generator (HFCG) coil with multi layer conductor wire filaments (MLCWF) in the form of rectangular cross-section. This approach is based on vespiary regular hexagonal (VRH) model. In this method, wire filaments of the generator coil are separated into many very small simi...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
دوره 17 3 شماره
صفحات -
تاریخ انتشار 2015