Rapid LC-APCI-MS-MS Method for Simultaneous Determination of Phenacetin and Its Metabolite Paracetamol in Rabbit Plasma
نویسندگان
چکیده
A highly sensitive liquid chromatographic-atmospheric pressure chemical ionization-tandem mass spectrometric method is developed to quantitate phenacetin and its metabolite paracetamol in rabbit plasma. The analytes and internal standard oxazepam are extracted from plasma by liquid–liquid extraction using ethyl acetate, and separated on a Zorbax SB-C18 column (2.1 mm 9 150 mm, 5 lm) using acetonitrile–0.1% formic acid in water (40:60 v/v) at a flow of 0.4 mL min. Detection is carried out by multiple reaction monitoring on a iontrap LC-MS-MS system with an atmospheric pressure chemical ionization interface. The assay is linear over the range 4–1,600 ng mL for phenacetin and 3–2,000 ng mL for paracetamol, with a lower limit of quantitation of 4 ng mL for phenacetin and 3 ng mL for paracetamol. Intraand inter-day precision are less than 7.1% and the accuracy are in the range 97.3–103.5%. The validated method is successfully used to analyze the drug in samples of rabbit plasma for pharmacokinetic study.
منابع مشابه
Method development for simultaneous determination of 41 pesticides in rice using LC-MS/MS technique and its application for the analysis of 60 rice samples collected from Tehran market
For the first time, a multi-residue methodfor simultaneous determination of 41LC-amenable pesticides in rice, belonging to different chemical classes has been developed in Iran by LC-MS/MS. The pesticides were analyzed simultaneously in a single run using positive electrospray ionization with multiple reaction monitoring (MRM) after extraction with slightly modified QuEChERS method. The calibra...
متن کاملMethod development for simultaneous determination of 41 pesticides in rice using LC-MS/MS technique and its application for the analysis of 60 rice samples collected from Tehran market
For the first time, a multi-residue methodfor simultaneous determination of 41LC-amenable pesticides in rice, belonging to different chemical classes has been developed in Iran by LC-MS/MS. The pesticides were analyzed simultaneously in a single run using positive electrospray ionization with multiple reaction monitoring (MRM) after extraction with slightly modified QuEChERS method. The calibra...
متن کاملComparison of ESI– and APCI–LC–MS/MS methods: A case study of levonorgestrel in human plasma☆
Electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI) techniques for liquid chromatography-tandem mass spectrometry (LC-MS/MS) determination of levonorgestrel were evaluated. In consideration of difference in ionization mechanism, the two ionization sources were compared in terms of LC conditions, MS parameters and performance of method. The sensitivity for detection...
متن کاملDevelopment and Validation of Bioanalytical Method for Simultaneous Estimation of Nebivolol Enantiomers in Human Plasma Using Liquid Chromatography-tandem Mass Spectrometry
The present study describes a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the simultaneous determination of S-RRR and R-SSS nebivolol (nebivolol enantiomers) in human plasma using solid phase extraction technique. Method of both S-RRR and R-SSS nebivolol (nebivolol enantiomers) has been developed and validated using racemic nebivolol D4 as an internal standard. Analytes...
متن کاملA quick and Sensitive Liquid Chromatography–tandem Mass Spectrometry (LC-MS) Method for the Determination of Enalapril and Enalaprilat in Human Plasma: Application to a Bioequivalence Study
A rapid and sensitive liquid chromatography–tandem mass spectrometry (LC-MS) method for the estimation of enalapril and enalaprilat in human plasma. Detection of analytes was achieved by tandem mass spectrometry with electrospray ionization (ESI) interface in positive ion mode was operated under the multiple-reaction monitoring mode. Sample pretreatment involved in a one-step protein precipitat...
متن کامل